background image

HND-T205 

page 16 

 

HND-T205 K05/0422 

 

4-Wire-Measuring 

When using resistance thermometers as the Pt100 a quite large measuring error 
can  be  caused  by  inadequate  cables  and  connections.  Using  4wire  measuring 
avoids  these  kinds  of  errors  mainly  caused  by  unwanted  resistances.  It  is 
suggested  to  use  suitable  probes  and  extensions  only.  (For  pin  assignment 
please refer to chapter 

7.7 Probe connection

 

 

Heat loss caused by probe construction:  

Especially  when  measuring  temperatures  which  deviate  very  much  from  the 
ambient temperature, measuring errors often occur if the heat loss caused by the 
probe  is  not  considered.  When  measuring  fluids  therefore  the  probe  should  be 
emerged  sufficiently  deep  and  be  stirred  continuously.  When  measuring  gases, 
the  probe  should  also  emerge  as  deep  as  possible  in  the  gas  to  be  measured 
(e.g. when measuring in channel/pipes) and the gas should flow around the probe 
at sufficient flow. 

 

 

Measuring Surface Temperature  

If  temperature  of  the  surface  of  an  object  has  to  be  measured,  one  should  pay 
attention especially when measuring hot (or very cold) surfaces, that the ambient 
air cools (or heats) the surface. Additionally, the object will be cooled (or heated) 
by the probe or the probe can have a better heat flow to the ambient temperature 
as to the objects surface.  
Therefore,  specially  designed  surface  probes  should  be  used.  The  measuring 
precision depends mainly on the construction of the probe and of the physics of 
the surface itself. If selecting a probe try to choose one with low mass and heat 
flow  from  sensor  to  handle.  Thermally  conductive  paste  can  increase  the 
precision in some cases. 
 

 

Allowable temperature Range Of Probes  

Pt100 Sensors are defined over a wide temperature range. Depending on probe 
materials  and  sort  of  sensor  (e.g.  hybrid  sensors,  wire  wound  resistors...)  the 
allowable  temperature  ranges  have  to  be  considered.  Exceeding  the  ranges  at 
least  causes  a  wrong  measuring,  it  may  even  damage  the  probe  permanently! 
Often it also has to be considered, that the temperature range is just valid for the 
probe tube, (plastic-) handles can’t stand the same high temperatures. Therefore, 
the  tube  length  should  be  selected  long  enough,  that  temperature  keeps  low  at 
the handle. 

 

Self Heating 

The  measuring  current  of  the  instrument  is  just  0.3  mA.  Because  of  this 
comparably low current practically now self heating effect has to be considered, 
even at air with low movement the self heating is <= 0.01 °C. 
 

 

Cooling by Evaporation 

When  measuring  air  temperature,  the  probe  has  to  be  dry.  Otherwise,  the  cold 
due to the evaporation causes too low measurings. 
 

Содержание HND-T205

Страница 1: ...Operating Instructions for Precision Hand Held Thermometer Model HND T205...

Страница 2: ...6 Special Functions 9 7 7 Probe connection 12 7 8 Operation of Logger 13 7 9 Some Basics of Precision Temperature Measuring 15 7 10 Fault and System Messages 17 7 11 Calibration Services 17 8 Maintena...

Страница 3: ...nly to be used maintained and serviced by persons familiar with these operating instructions and in accordance with local regulations applying to Health Safety and prevention of accidents When used in...

Страница 4: ...nd zero point increase entry for example 6 Electrical Connection 6 1 Mains Operation Attention When using a power supply unit please note that operating voltage has to be 10 5 to 12 VDC Do not apply o...

Страница 5: ...her devices Internal connection in third party devices e g connection GND and earth may result in not permissible voltages impairing or destroying the device or another device connected Warning Operat...

Страница 6: ...olated interface adapter accessory HND Z031 Operation as analogue output Connection via suitable cable Attention The output mode has to be configured p r t 7 6 6 and influences battery life 2 Probe co...

Страница 7: ...he min Value will be deleted Key 6 Measurement Hold current measuring value HLD in display or operation of logger functions p r t chapter 7 8 7 5 Device Configuration Note Some menu items will be show...

Страница 8: ...e will be automatically switched off as soon as this time has elapsed if no key is pressed no interface communication takes place oFF Power off function inactive continuous operation e g mains operati...

Страница 9: ...besides the standard conversion of resistance to temperature following EN60751 Lin E 751 also other curves can be used The user sensor curve can be read and edited by configuration software The standa...

Страница 10: ...ard curve LIN E 751 and the user sensor curve Lin USEr posses separate correction settings 7 6 5 Average filter tAuG The filter forms the arithmetic floating average over the entered time The displaye...

Страница 11: ...on 23 read max alarm rail AL AL Hi 216 read zero displacement 32 read configuration flag alarm function 1 Alarm horn 3 BitLoggerOn 50 BitCyclicLogger 51 217 set zero displacement 222 Read power off ti...

Страница 12: ...urned interface message 7 6 8 Real Time Clock CLOC The real time clock is used for the logger function Recorded values are also containing the point of time when they were measured Please check the se...

Страница 13: ...tware also allows easy configuration and starting of the logger When the logger is activated Func Stor or Func CYCL the hold function is no longer available key 6 is solely used for the operation of t...

Страница 14: ...hapter 8 Device configuration Max number of measurings 16384 a Starting a recording press for 2 seconds the recording will be started Each recording is signalled with a short display of St XXXXX XX is...

Страница 15: ...03 0 0005 temperature A 0 15 0 002 temperature Error over measuring range Error over range 50 150 C For applications demanding higher precision than given by these classes we suggest to adjust the dev...

Страница 16: ...probe or the probe can have a better heat flow to the ambient temperature as to the objects surface Therefore specially designed surface probes should be used The measuring precision depends mainly o...

Страница 17: ...to manufacturer for repair Sensor error no sensor connected Connect sensor to socket sensor cable or device defective return to manufacturer for repair Err 1 Value exceeding measuring range Check Is t...

Страница 18: ...99 99 199 99 C or 200 0 850 0 C Fahrenheit values accordingly Accuracy at nominal temperature 25 C 0 03 C in the range 199 99 199 99 C 0 1 C 1 digit in the range 200 0 850 0 C Resolution 0 01 C or 0 1...

Страница 19: ...0 min can be deactivated Zero point and increase entry zero point and increase correction can be entered digitally Logger functions Manual 99 datasets Cyclic 16 384 datasets Adjustable cycle time 1s 1...

Страница 20: ...50 400 C approx 10 sec HND TF02 Immersion probe for liquids and gases 4 wire Like HND TF01 but with 1 10DIN Class B 0 03 C at 0 C and flexible sheath tube 3 mm 50 400 C approx 10 sec HND TF03 Insertio...

Страница 21: ...gg crate foam 275x229x83 mm HND Z022 Large case with recess 394x294x106 mm HND Z023 Interface converter on RS232 galvanically isolated HND Z031 Interface converter on USB galvanically isolated HND Z03...

Страница 22: ...elow EN 61326 1 2013 Electrical equipment for measurement control and laboratory use EMC requirements Part 1 General requirements EN IEC 63000 2018 Technical documentation for the assessment of electr...

Отзывы: