HND-P215 

HND-P215 K06/0422 

 

page 3 

2.  Note 

Please read these operating instructions before unpacking and putting the unit into 
operation. Follow the instructions precisely as described herein.  
The  instruction  manuals  on  our  website  www.kobold.com  are  always  for  currently 
manufactured  version  of  our  products.  Due  to  technical  changes,  the  instruction 
manuals  available  online  may  not  always  correspond  to  the  product  version  you 
have  purchased.  If  you  need  an  instruction  manual  that  corresponds  to  the 
purchased  product  version,  you  can  request  it  from  us  free  of  charge  by  email 
([email protected])  in  PDF  format,  specifying  the  relevant  invoice  number  and 
serial  number.  If  you  wish,  the  operating  instructions  can  also  be  sent  to  you  by 
post in paper form against an applicable postage fee.

 

 
The devices are only to be used, maintained and serviced by persons familiar with 
these  operating  instructions  and  in  accordance  with  local  regulations  applying  to 
Health & Safety and prevention of accidents.  
 
When  used  in  machines,  the  measuring  unit  should  be  used  only  when  the 
machines fulfill the EG-machine guidelines. 
 
 
 

3.  Instrument Inspection 

Instruments are inspected before shipping and sent out in perfect condition.  
Should damage to a device be visible, we recommend a thorough inspection of the 
delivery  packaging.  In  case  of  damage,  please  inform  your  parcel  service  / 
forwarding  agent  immediately,  since  they  are  responsible  for  damages  during 
transit. 
 

Scope of delivery: 

The standard delivery includes: 

  Manual Pressure Measuring Devices with External and Integrated Pressure 

Sensors  

model: HND-P215 

 
 
 
 

4.  Regulation Use 

Any  use  of  the  Manual  Pressure  Measuring  Devices  with  External  and  Integrated 
Pressure  Sensors,  model:  HND-P215,  which  exceeds  the  manufacturer’s 
specification may invalidate its warranty. Therefore, any resulting damage is not the 
responsibility of the manufacturer. The user assumes all risk for such usage.  
 
 
 

Содержание HND-P215

Страница 1: ...Operating Instructions for Manual Pressure Measuring Devices with External and Integrated Pressure Sensors Model HND P215...

Страница 2: ...The Serial Interface 15 8 2 Analogue Output Scaling with DAC 0 and DAC 1 17 8 3 Instrument Adjustment 17 8 4 Pressure Connection to the Sensors 18 8 5 Relative Pressure Sensors Types HND PS01 PS05 HN...

Страница 3: ...familiar with these operating instructions and in accordance with local regulations applying to Health Safety and prevention of accidents When used in machines the measuring unit should be used only...

Страница 4: ...lue memory hold function automatic self shut off or zero point offset entry for all connected pressure sensors for example The HND P215 type also has a logger function a peak value memory or a minimum...

Страница 5: ...e device to other devices the connection has to be designed most thoroughly as internal connections in third party devices e g connection GND with protective earth may lead to undesired voltage potent...

Страница 6: ...bel open again to swing open the pop up clip further Function The device with a closed pop up clip can be plainly laid onto a table or attached to a belt etc The device with pop up clip at position 90...

Страница 7: ...and device is operated under steady conditions a permanent zero point adjustment can be carried out To carry out the adjustment press button 3 for approx 5 seconds Auto Null will be displayed shortly...

Страница 8: ...The parameters can be changed with key 2 or key 5 Pressing Menu again jumps back to the main configuration menu and saves the settings Quit key 6 finishes the configuration and returns to standard me...

Страница 9: ...obe or measuring device oFF Scale correction factor inactive 0 000 SEt AL Set Alarm Settings Of Alarm Function AL 1 On Alarm sensor 1 on with buzzer sound no So Alarm sensor 1 on without buzzer sound...

Страница 10: ...lic logger function records the arithmetic mean value the highest and the lowest peak of the refering time interval Attention higher power consumption measuring is sensitive to noise EMI 7 7 1 3 rAtE...

Страница 11: ...ng a selectable period of time and then calculates the average display value It is independent from the selected kind of measuring slow fast peak detect As long as not enough values are collected sele...

Страница 12: ...on logger setting min peak and max peak of sensor 1 sensor 2 Min and max peak are the minimum resp the maximum of the measured values since the last recording Using them allows f e analysis of fluctua...

Страница 13: ...ress 2 seconds Set key 4 The first menu displayed now is rEAd LoGG read logger data After pressing key 3 the measurement recorded last will be displayed changing between the different values refering...

Страница 14: ...r 1 sensor 2 min peak max peak since the last recording Starting a recording By pressing Store key 6 for 2 seconds the recording will be initiated After that the display shows St XXXX for a short time...

Страница 15: ...log output 0 1V If none of both is needed we suggest to switch the output off because battery life then is extended 8 1 The Serial Interface By means of the serial interface and a suitable electricall...

Страница 16: ...ime Conf P oFF x 12 Read ID number x 223 Set power off time Conf P oFF x 22 Read min alarm rail AL AL Lo x x x 224 Logger Read data of CYCL Logger x 23 Read max alarm rail AL AL Hi x 225 Logger Read c...

Страница 17: ...ightly above 1V to the output plug wiring 8 3 Instrument Adjustment 8 3 1 Zero Displacement Sensor 1 OFS 1 and Sensor 2 OFS 2 A zero displacement can be carried out for the measured value value displa...

Страница 18: ...e port A Please note that all values are displayed as positive values No minus sign will be shown Example for HND PS02 For tube connection B the measuring range covers 19 99 to 25 00 mbar If you replu...

Страница 19: ...r supply replace it when necessary System error Disconnect battery and power supplies wait shortly then reconnect Device defective Return to manufacturer for repair Err 1 Measured value above allowabl...

Страница 20: ...ing device above 50 C We recommend taking out battery if device is not used for a longer period of time After recommissioning the real time clock has to be set again 10 Technical Information Measureme...

Страница 21: ...ing rate 7 0 mA Low Power Logger 0 3 mA for cycle time 30s without interface communication active and no alarm horn sounding up to 0 8 mA at cycle time 1s Low battery warning bat Working conditions 20...

Страница 22: ...14 0 70 00 bar abs 0 1 bar max 280 bar abs HND PS15 0 160 0 bar abs max 600 bar abs HND PS16 0 250 0 bar abs HND PS17 0 400 0 bar abs HND PS18 0 400 mbar rel 0 1 mbar max 2 bar rel 0 70 C G male threa...

Страница 23: ...anically isolated HND Z032 Interface converter on USB galvanically isolated HND Z033 Adapter RS232 converter on USB interface HND Z034 Windows software for setting and data read and print out of instr...

Страница 24: ...tandards noted below EN 61326 1 2013 Electrical equipment for measurement control and laboratory use EMC requirements Part 1 General requirements EN IEC 63000 2018 Technical documentation for the asse...