![Kobold HND-P215 Скачать руководство пользователя страница 19](http://html1.mh-extra.com/html/kobold/hnd-p215/hnd-p215_operating-instructions-manual_1982446019.webp)
HND-P215
HND-P215 K06/0422
page 19
8.6 Error and System Messages
Display
Meaning
What to do?
Low battery power, device will only
continue operation for a short period of
time
Replace battery
Battery empty
Replace battery
Mains operation without battery: wrong
voltage
Check power supply, replace it when
necessary
or
Err.9
No sensor connected
Switch off device and connect sensor
Connected sensor or device defective
If 2nd sensor available, check if device is ok.
Return defective device/sensor to manufacturer
for repair
Value extremely out of measuring range
Check: pressure not within sensor range?
No display or
confused
characters,
device does not
react on
keypress
Battery empty
Replace battery
Mains operation: wrong voltage or polarity Check power supply, replace it when
necessary
System error
Disconnect battery and power supplies, wait
shortly, then reconnect
Device defective
Return to manufacturer for repair
Err.1
Measured value above allowable range
Check: pressure not within sensor range?
-> measuring value to high!
Sensor defective
Return to manufacturer for repair
Err.2
Measured value below allowable range
Check: pressure not within sensor range?
-> measuring value to low!
Sensor defective
Return to manufacturer for repair
Err.3
Display range overflow
Check: value above 19999 -> to high to be
displayed!
Err.4
Display range underflow
Check: value below 19999 (Tara?) -> to low!
Er.11
Value could not be calculated
Choose different unit
Calculation overflow happened
Choose different unit
Err.7
System error
Return to manufacturer for repair
----
Sensor not present/recognized
Reconnect sensor,
During logging: stop the logger and restart it
Could not calculate value
Suitable sensor/unit combination necessary