Kobold HND-P215 Скачать руководство пользователя страница 18

HND-P215 

page 18 

 

HND-P215 K06/0422 

8.4  Pressure Connection to the Sensors 

The  device  is  designed  to  be  connected  to  the  sensors  of  the  HND-PS...-series 
without a new calibration being necessary. Therefore, a great variety of replaceable 
sensors  of  e.g.  –1.999...2.500  mbar  relative  up  to  0...1000  bar  absolute  pressure 
can be connected to the device. 
 
 

8.5  Relative Pressure Sensors (Types: HND-PS01…-PS05, HND-

PS09) 

 

For measurements of over- or under pressure: 

Connect plastic tube with internal dia of 4 mm to pressure port "B". Port "A" is not 
used! 
 

 

For  measurements  of  under  pressure:

  (with  higher  negative  measuring 

range)  

Connect plastic tube with internal dia of 4 mm to pressure port "B". Port "A" will not 
be used!  
Pressure sensors HND-PS01, HND-PS02 and HND-PS03 allow for measurements 
of  under  pressure  up  to  the  entire  over  pressure  measuring  range  by  re-plugging 
the tube to pressure port "A". Please note that all values are displayed as positive 
values. No minus sign will be shown. (Example for HND-PS02: For tube connection 
"B"  the  measuring  range  covers  -19.99  to  25.00  mbar.  If  you  replug  to  port  "A" 
under pressure measurements down to -25.00 mbar could be carried out with the 
display showing the value 25.00 (no minus sign). 
 

Note: All values are displayed now as positive values. No minus sign will be 
shown. Example: it is possible to measure under pressure down to -
25.00 mbar, the display shows then the value 25.00 (no minus sign). 
 

For measurements of differential pressure: 

Connect  both  plastic  tubes  with  an  internal  dia  of  4  mm  to  pressure  port  "B"  and 
"A"; make sure to apply higher pressure to port "B". 
 

Stainless steel pressure sensors: (types: HND-PS01…-PS30) 

For  measurements  of  over-,  under-  or  absolute  pressure  screw  sensor  to  G1/4" 
pressure terminal or plug plastic tube to a suitable adapter. 
 

Measurements of differential pressure with two sensors  

By  means  of  the  calculation  sensor  1  –  sensor  2  (DIF)  press.  differences  of  any 
sensor combinations can be measured. 
 
 

Содержание HND-P215

Страница 1: ...Operating Instructions for Manual Pressure Measuring Devices with External and Integrated Pressure Sensors Model HND P215...

Страница 2: ...The Serial Interface 15 8 2 Analogue Output Scaling with DAC 0 and DAC 1 17 8 3 Instrument Adjustment 17 8 4 Pressure Connection to the Sensors 18 8 5 Relative Pressure Sensors Types HND PS01 PS05 HN...

Страница 3: ...familiar with these operating instructions and in accordance with local regulations applying to Health Safety and prevention of accidents When used in machines the measuring unit should be used only...

Страница 4: ...lue memory hold function automatic self shut off or zero point offset entry for all connected pressure sensors for example The HND P215 type also has a logger function a peak value memory or a minimum...

Страница 5: ...e device to other devices the connection has to be designed most thoroughly as internal connections in third party devices e g connection GND with protective earth may lead to undesired voltage potent...

Страница 6: ...bel open again to swing open the pop up clip further Function The device with a closed pop up clip can be plainly laid onto a table or attached to a belt etc The device with pop up clip at position 90...

Страница 7: ...and device is operated under steady conditions a permanent zero point adjustment can be carried out To carry out the adjustment press button 3 for approx 5 seconds Auto Null will be displayed shortly...

Страница 8: ...The parameters can be changed with key 2 or key 5 Pressing Menu again jumps back to the main configuration menu and saves the settings Quit key 6 finishes the configuration and returns to standard me...

Страница 9: ...obe or measuring device oFF Scale correction factor inactive 0 000 SEt AL Set Alarm Settings Of Alarm Function AL 1 On Alarm sensor 1 on with buzzer sound no So Alarm sensor 1 on without buzzer sound...

Страница 10: ...lic logger function records the arithmetic mean value the highest and the lowest peak of the refering time interval Attention higher power consumption measuring is sensitive to noise EMI 7 7 1 3 rAtE...

Страница 11: ...ng a selectable period of time and then calculates the average display value It is independent from the selected kind of measuring slow fast peak detect As long as not enough values are collected sele...

Страница 12: ...on logger setting min peak and max peak of sensor 1 sensor 2 Min and max peak are the minimum resp the maximum of the measured values since the last recording Using them allows f e analysis of fluctua...

Страница 13: ...ress 2 seconds Set key 4 The first menu displayed now is rEAd LoGG read logger data After pressing key 3 the measurement recorded last will be displayed changing between the different values refering...

Страница 14: ...r 1 sensor 2 min peak max peak since the last recording Starting a recording By pressing Store key 6 for 2 seconds the recording will be initiated After that the display shows St XXXX for a short time...

Страница 15: ...log output 0 1V If none of both is needed we suggest to switch the output off because battery life then is extended 8 1 The Serial Interface By means of the serial interface and a suitable electricall...

Страница 16: ...ime Conf P oFF x 12 Read ID number x 223 Set power off time Conf P oFF x 22 Read min alarm rail AL AL Lo x x x 224 Logger Read data of CYCL Logger x 23 Read max alarm rail AL AL Hi x 225 Logger Read c...

Страница 17: ...ightly above 1V to the output plug wiring 8 3 Instrument Adjustment 8 3 1 Zero Displacement Sensor 1 OFS 1 and Sensor 2 OFS 2 A zero displacement can be carried out for the measured value value displa...

Страница 18: ...e port A Please note that all values are displayed as positive values No minus sign will be shown Example for HND PS02 For tube connection B the measuring range covers 19 99 to 25 00 mbar If you replu...

Страница 19: ...r supply replace it when necessary System error Disconnect battery and power supplies wait shortly then reconnect Device defective Return to manufacturer for repair Err 1 Measured value above allowabl...

Страница 20: ...ing device above 50 C We recommend taking out battery if device is not used for a longer period of time After recommissioning the real time clock has to be set again 10 Technical Information Measureme...

Страница 21: ...ing rate 7 0 mA Low Power Logger 0 3 mA for cycle time 30s without interface communication active and no alarm horn sounding up to 0 8 mA at cycle time 1s Low battery warning bat Working conditions 20...

Страница 22: ...14 0 70 00 bar abs 0 1 bar max 280 bar abs HND PS15 0 160 0 bar abs max 600 bar abs HND PS16 0 250 0 bar abs HND PS17 0 400 0 bar abs HND PS18 0 400 mbar rel 0 1 mbar max 2 bar rel 0 70 C G male threa...

Страница 23: ...anically isolated HND Z032 Interface converter on USB galvanically isolated HND Z033 Adapter RS232 converter on USB interface HND Z034 Windows software for setting and data read and print out of instr...

Страница 24: ...tandards noted below EN 61326 1 2013 Electrical equipment for measurement control and laboratory use EMC requirements Part 1 General requirements EN IEC 63000 2018 Technical documentation for the asse...

Отзывы: