7.4 Push buttons
On/off key
in/max when taking measurements:
press shortly:
min. or max. measuring value will
be displayed
press for 1 sec.: the value shown will be deleted
up/down for configuration:
to enter values, and/or change settings
CAL:
(for HND-FF31-measuring probe only)
press for 2 sec.: humidity calibration will be started
press for more than 10 sec.: reset of humidity
calibration to factory
calibration
Set/Menu:
press (Set) shortly: display changes between:
T1,T2,Td,
∆
Td,kJ/kg (if existing)
press (Menu) for 2 sec.: configuration menu is
activated
Store/Quit:
Measurement: Hold current measuring value ('HLD'
in display)
for flow measurements in the 'AVGHold' mode:
start new measurement or handling
of logger functions
Set/Menu:
Acknowledge setting, return to
measuring.