7.9  The Serial Interface 

All  measuring  and  setting  data  of  the  device  can  be  read  and/or  changed  by 
means of the serial interface and a suitable electrically isolated interface adapter 
(HND-Z031, HND-Z032). In order to avoid transmission errors, there are several 
security checks implemented. 
The following 

standard software packages

 are available for data transfer: 

-

 

HND-Z034

:  

Software for temperature display and/or read out of logger 
data. 

-

 

BUS-S20M: 

20-channel software to display the humidity (channel 1), the 
temperature. (channel 2, 3) 

 

The following interface functions will be supported:

 

Channel 

DLL- 
Code 

Name/function 

6

x

Read nominal value

x

Read system status

 

 

 

 

 

12 

Read ID number

1) 

1) 

1) 

1) 

1) 

1)

22 

Read min. alarm limit

1) 

1) 

1) 

1) 

1) 

1)

23 

Read max. alarm limit

2) 

 

 

 

 

 

32 

Read configuration flag

2) 

 

 

 

 

 

160 

Set configuration flag

x

199 

Read meas. type in display

x

200 

Read min. display range

x

201 

Read max. display range

x

202 

Read unit of display

x

204 

Read decimal point of display

 

 

 

 

 

208 

Read channel count

 

 

 

 

216 

Read offset correction

 

 

 

 

 

218 

Read corr. factor (1000..1200)

3) 

3) 

3) 

3) 

3) 

3)

224 

Read logger data (cyclic logger)

4) 

 

 

 

 

 

225 

Read logger cycle

5) 

 

 

 

 

 

226 

Set logger cycle

6) 

 

 

 

 

 

227 

Start logger recording

7) 

 

 

 

 

 

228 

Read count of logger data

7) 

 

 

 

 

 

229 

Read logger state

3) 

 

 

 

 

 

231 

Read logger stop time

x) 

 

 

 

 

 

233 

Read real-time clock

 

 

 

 

 

234 

Set real-time clock

7) 

 

 

 

 

 

236 

Read logger size

 

 

 

 

 

240 

Reset unit

 

 

 

 

 

254 

Read program identification

8) 

 

 

 

 

 

260 

Read logger data (man. logger)

1)  only when alarm is activated for referring channel 
2)  configuration flags:   50: 0 = logger off 

1 = logger on 

 

 

51: 0 = man. logger  1 = cyclic logger 

3)  only when logger function = CYCL, data present and logger stopped. 
4)  only when logger function = CYCL 
5)  only when logger function = CYCL and no data in memory 
6)  only when logger function = Stor, or logger function = CYCL and no data in memory 
7)  only when logger is activated (CYCL or Stor) 
8)  only when logger function = Stor and data in memory 

For HND-FF31 
 Channel 1: rel atmospheric humidity 
 Channel 2: temperature T1 
 Channel 3: temperature T2 
 Channel 4: dew point temperature Td 
 Channel 5: dew point distance 

Td 

 Channel 6: enthalpy h 
 
For HND-FF32/HND-FF33 
 Channel 1: flow speed 
 Channel 3: temperature T2 
Channel 2, 4, 5, 6: not supported. 
 
For NiCr-Ni 

(without HND-FF31/32/33) 

 Channel 3: temperature T2 
Channel 1, 2, 4, 5, 6: not supported. 
Logger handling still works with 
channel 1. 

Содержание HND-F215

Страница 1: ...Operating Instructions for Manual Humidity Precision Measuring Unit Model HND F215...

Страница 2: ...tes for Special Functions 13 7 9 The Serial Interface 17 7 10 How to Calibrate Meas of Rel Humidity Using HND FF31 18 7 11 System and Error Messages 20 7 12 System and Error Messages 21 7 13 Error and...

Страница 3: ...used maintained and serviced by persons familiar with these operating instructions and in accordance with local regulations applying to Health Safety and prevention of accidents When used in machines...

Страница 4: ...ng voltage has to be 10 5 to 12 VDC Do not apply over voltage Cheap 12 V power supply devices often have excessive no load voltage We therefore recommend using regulated voltage power supply devices T...

Страница 5: ...y has to be designed most carefully Internal connection in third party devices e g connection GND and earth may result in not permissible voltages impairing or destroying the device or another device...

Страница 6: ...F32 flow speed water 0 05 5m s 7 3 Displays Depending on the measuring probes sensors connected the following measuring results can be displayed HND FF31 Main display r H relative atmospheric humidity...

Страница 7: ...arrow indicates that correction factor is activated Logger arrow indicates that the logger function is activated Alarm arrow indicates an alarm Messages at device startup The device will show some mes...

Страница 8: ...g probe only press for 2 sec humidity calibration will be started press for more than 10 sec reset of humidity calibration to factory calibration Set Menu press Set shortly display changes between T1...

Страница 9: ...s marked by 1 will only be displayed if no data is stored in the logger AVG Selection of Averaging Proceedings for Flow Measurement 1 only HND FF32 33 Cont continuous averaging the average value calcu...

Страница 10: ...the base addresses accordingly Set Alarm Alarm Settings Settings for the alarm function Please note the points marked by 2 will only be displayed if the alarm functions on or no So have been selected...

Страница 11: ...The scope of supply includes one sensor to measure relative atmospheric humidity and another one to measure the ambient temperature T1 rel humidity r H Relative humidity measured in the tip of the pro...

Страница 12: ...T1 T2 and relative atmospheric humidity The combination sensor is used to measure the ambient air whose condition is issued to calculate the dew point Td The surface sensor is used to measure surface...

Страница 13: ...s the HND F215 instrument has been switched on the device starts calculating the average flow value during the averaging time During measuring the current measuring value will be shown in the top line...

Страница 14: ...ls 2 and 3 will have the following addresses Example base address 21 channel 1 21 channel 2 22 channel 3 23 Alarm 3 alarm settings are available off off on with horn sound on on no horn sound no So De...

Страница 15: ...d time date Press Store key to store current values St XX will be displayed for a short time XX representing the number of the data set 1 99 If the logger memory is full a warning will appear on the d...

Страница 16: ...g value Stop logger recording Press Store key key 6 for a short time to stop recording You will then be asked to acknowledge again recording to be stopped recording to be continued Use the keys key 2...

Страница 17: ...Read offset correction x 218 Read corr factor 1000 1200 3 3 3 3 3 3 224 Read logger data cyclic logger 4 225 Read logger cycle 5 226 Set logger cycle 6 227 Start logger recording 7 228 Read count of l...

Страница 18: ...libration equipment is quite often highly dependent on temperature This dependence is automatically compensated for when calibrating with the integrated calibration equipment and automatic detection I...

Страница 19: ...detected humidity value measured may deviate from value set by manufacturer by approx 10 In case of manual input enter value here As soon as the display stops blinking and changing between values a st...

Страница 20: ...OC xx xx identified sensor HND FF31 FF32 or FF33 temperature offset of the HND FF31 sensor only with HND FF31 and adj offset value off flow averaging procedure only with HND FF32 FF33 AVG Hold or AVG...

Страница 21: ...mains op power supply defective or wrong voltage polarity check replace power supply System error disconnect battery or power supply wait for a short time re connect device defective return to manufa...

Страница 22: ...the battery has been used up and needs to be replaced The device will however operate correctly for a certain time If bAt is shown in the upper display the voltage is too low to operate the device th...

Страница 23: ...onal Power supply 9 V monobloc battery included in the scope of delivery external 10 5 12 VDC via jack Material housing made of impact resistant ABS plastic Protection IP 65 front Dimensions 142 x 71...

Страница 24: ...sor head and tube 40 60 C short term 120 C Dimensions probe 14 x 119 mm handle 19 x 135 mm 1 m cable HND FF32 Flow meas probe with spring head calibrated and completely replaceable Measuring range 0 0...

Страница 25: ...rter on USB interface HND Z034 Windows software for setting data read out and printing of the data of housings of the HND series with logger function HND Z052 100 ml conductivity paste for surface mea...

Страница 26: ...low EN 61326 1 2013 Electrical equipment for measurement control and laboratory use EMC requirements Part 1 General requirements EN 50581 2012 Technical documentation for the assessment of electrical...

Страница 27: ...n Hard Acronychia laevis h 28 5 66 Ayan Distemonanthus benthamianus h 285 7 54 Balau Shorea laevis h 31 4 54 Balau red Shorea guiso h 32 4 68 Balsa Ochroma pyramidale h 33 4 91 Basralocus Angelique Di...

Страница 28: ...Cypress Northern Callitris intratropica h 115 6 78 Cypress Rottnest Island Callitris preisii h 116 7 80 Cypress White Callitris glaucophylla h 117 6 86 Dakua Salusalu Fiji Decussocarpus vitiensis h 1...

Страница 29: ...leri h 243 8 73 Mahogany New Guinea Dysoxylum spp h 241 6 74 Mahogany Red Eucalyptus botryoides h 244 7 91 Mahogany Rose Dysoxylum fraseranum h 245 7 65 Mahogany Southern Eucalyptus botryoides h 246 5...

Страница 30: ...e Celery Top Phyllocladus aspenifolius h 329 7 71 Pine Hoop Araucaria cunninghamii h 330 7 79 Pine Huon Dacrydium franklinii h 331 8 70 Pine King William Athrotaxis selaginoides h 332 7 67 Pine Klinki...

Страница 31: ...atopetalum succirubrum h 408 7 63 Tallowwood Eucalyptus microcorsis h 409 4 92 Tatajuba Bagassa guianesis h 30 7 44 Taun Maleisien Pometia pinnata h 195 0 105 Taun New Guinea Pometia pinnata h 196 6 1...

Страница 32: ...7 yes Plaster synthetic b 12 18 2 60 8 yes On wall plaster b 20 0 0 38 8 no Lime mortar KM 1 3 b 14 0 4 40 4 yes Lime sand bricks 14 DF 200 gross density 1 9 b 28 0 1 12 5 yes Limestone b 15 0 4 29 5...

Отзывы: