![Knick Stratos Pro A401B PH Скачать руководство пользователя страница 99](http://html.mh-extra.com/html/knick/stratos-pro-a401b-ph/stratos-pro-a401b-ph_user-manual_1980199099.webp)
99
Configuration
Menu item
Action
Choices
Use of relays
Select in the text line
using
keys:
• Limit function (LIMITS)
• Controller
(CONTROLLER)
Press enter to confirm.
Note: Selecting
CONTROLLER leads to
Controller menu group
CTR.
Select process
variable
Select desired process
variable using
keys.
Press enter to confirm.
Limit 1 function
Select desired function
using
keys.
Lo Level: active if value
falls below setpoint /
Hi Level: active if value
exceeds setpoint
Press enter to confirm.
Limit 1 icon:
Limit 1
contact response
N/O: normally open
contact
N/C: normally closed
contact
Select using
keys.
Press enter to confirm.
Limit 1 setpoint
Enter setpoint using
keys.
Press enter to confirm.