Configuration
Menu item
Action
Choices
Time averaging filter
Enter value using
keys.
Press enter to confirm.
Time averaging filter
To smoothen the current output, a low-pass filter with adjustable filter
time constant can be switched on. When there is a jump at the input
(100 %), the output level is at 63 % after the time interval has been
reached. The time interval can be set from 0 to 120 sec. If the time
interval is set to 0 sec, the current output directly follows the input.
Please note:
The filter only acts on the current output, not on the display, the limit
values, or the controller!
During HOLD the filter is not applied. This prevents a jump at the
output.
Time interval 0...120 s
5
pH
0/4-20 mA
Time interval 0...120 s
Temp
Display / Controller