45
Message
Released by
Cause
Alarm
Sensocheck
Polarization / Cable
(22 mA)
Error messages
Flow (CONTROL input)
Alarm
contact
opens
ERR 10: ORP display range violation
< -1999 mV or > 1999 mV
HOLD
HOLD
HOLD via menu or input
(Last/Fix)
CONF
Configuration
CAL
Calibration
SERVICE
Service
Alarm and HOLD Messages
Power supply
12...24 V AC/DC
13
12
Generating a message via the CONTROL input
(min. flow / max. flow)
The CONTROL input can be used for parameter set selection or for
flow measurement (pulse principle), depending on its assignment in
the “Configuration” menu.
When preset to flow measurement
an alarm can be generated when the measured flow exceeds a
specified range:
(specify value, default: 5 liters/h)
(specify value, default: 25 liters/h)
CONTROL
input