137
Diagnostics
In the Diagnostics mode you can access the following menus without
interrupting the measurement:
CALDATA
Viewing the calibration data
SENSOR
Viewing the sensor data
SELFTEST
Starting a device self-test
LOGBOOK
Viewing the logbook entries
MONITOR
Displaying currently measured values
VERSION
Displaying device type, software version, serial number
Access to diagnostics can be protected with a passcode
(SERVICE menu).
Please note:
HOLD is not active during Diagnostics mode!
Action
Key
Remark
Activate
Diagnostics
Press menu key to call the selection
menu.
(Display color changes to turquoise.)
Select DIAG using
keys,
confirm by pressing enter.
Select
diagnostics
option
Use
keys to select from:
CALDATA SENSOR SELFTEST
LOGBOOK MONITOR VERSION
See next pages for further
proceeding.
Exit
meas
Exit by pressing meas.
Diagnostics (DIAG)