35
Oxy
DATA INPUT Calibration
(Calibration by entering known sensor values)
The calibration method is selected in the configuration menu.
Measurement
cal
CAL
DATA INPUT
xx blinks
nA
ZERO POINT
Use
to set the known value for the sensor
zero point.
cal
xxx blinks
nA
SLOPE
Use
to set the known value for the sensor
slope.
cal
Calibration is performed. The meter then automatically returns to measuring
mode.
Note:
To abort calibration, you can press
meas
at any time.
Oxygen Calibration
Содержание Portavo 904X COND
Страница 2: ......
Страница 67: ...67 pH Buffer Tables 01 Mettler Toledo Nominal values in bold C pH ...
Страница 68: ...68 pH Buffer Tables 02 Knick CaliMat Nominal values in bold C pH ...
Страница 69: ...69 pH Buffer Tables 03 Ciba 94 Nominal values 2 06 4 00 7 00 10 00 C pH 1 extrapolated ...
Страница 70: ...70 pH Buffer Tables 04 Technical Buffers to NIST Nominal values in bold C pH 1 values added ...
Страница 72: ...72 pH Buffer Tables 06 HACH Nominal values 4 01 7 00 10 01 0 02 at 25 C C pH ...
Страница 73: ...73 pH Buffer Tables 07 WTW Technical Buffers Nominal values in bold C pH ...
Страница 74: ...74 pH Buffer Tables 08 Hamilton Nominal values in bold C pH ...
Страница 75: ...75 pH Buffer Tables 09 Reagecon Nominal values in bold C pH 1 values added ...
Страница 76: ...76 pH Buffer Tables 10 DIN 19267 Nominal values in bold C pH 1 extrapolated ...