20
pH
Measurement
cal
CAL
MANUAL
The number of calibration points has been
selected in the configuration menu.
CAL
1
/2/3
PRESS CAL
cal
pH display blinks
PRESS CAL
Take the temperature‑corrected pH value from
the buffer description and enter it using
.
cal
mV display blinks
Depending on the number of calibration points,
the procedure described above for CAL
1
/2/3 is
repeated.
mV value blinks until calibration is completed, then successive display of:
CAL DATA
1
/2/3 CAL POINTS
ZERO POINT
SLOPE
The meter then automatically returns to measuring mode.
MANUAL Calibration
(Manual calibration)
The calibration method is selected in the configuration menu.
Note:
To abort calibration, you can press
meas
at any time. This will be confirmed
by the “CAL ABORTED” display message. Exception: When you have selected “CAL
POINTS 1‑2‑3” and the first calibration step has been completed, the calibration pro‑
cess cannot be stopped any more.
pH Calibration
Содержание Portavo 904X COND
Страница 2: ......
Страница 67: ...67 pH Buffer Tables 01 Mettler Toledo Nominal values in bold C pH ...
Страница 68: ...68 pH Buffer Tables 02 Knick CaliMat Nominal values in bold C pH ...
Страница 69: ...69 pH Buffer Tables 03 Ciba 94 Nominal values 2 06 4 00 7 00 10 00 C pH 1 extrapolated ...
Страница 70: ...70 pH Buffer Tables 04 Technical Buffers to NIST Nominal values in bold C pH 1 values added ...
Страница 72: ...72 pH Buffer Tables 06 HACH Nominal values 4 01 7 00 10 01 0 02 at 25 C C pH ...
Страница 73: ...73 pH Buffer Tables 07 WTW Technical Buffers Nominal values in bold C pH ...
Страница 74: ...74 pH Buffer Tables 08 Hamilton Nominal values in bold C pH ...
Страница 75: ...75 pH Buffer Tables 09 Reagecon Nominal values in bold C pH 1 values added ...
Страница 76: ...76 pH Buffer Tables 10 DIN 19267 Nominal values in bold C pH 1 extrapolated ...