24
pH
ORP OFFSET Calibration
(available if an ORP or
pH/ORP combo sensor is connected)
Selected in the configuration menu.
pH Calibration
Measurement
cal
CAL
ORP OFFSET
You can specify an offset for the ORP value mea‑
sured by the sensor.
After calibration has been activated, the follow‑
ing values are listed in the display:
• ORP setpoint (in mV)
• temperature measured by sensor
• measured ORP value (in mV)
cal
“ORP setpoint” blinks
Use
to set the ORP value.
cal
Calibration is performed, the offset value is indicated.
The meter then automatically returns to measuring mode.
Note:
To abort calibration, you can press
meas
at any time.
Содержание Portavo 904X COND
Страница 2: ......
Страница 67: ...67 pH Buffer Tables 01 Mettler Toledo Nominal values in bold C pH ...
Страница 68: ...68 pH Buffer Tables 02 Knick CaliMat Nominal values in bold C pH ...
Страница 69: ...69 pH Buffer Tables 03 Ciba 94 Nominal values 2 06 4 00 7 00 10 00 C pH 1 extrapolated ...
Страница 70: ...70 pH Buffer Tables 04 Technical Buffers to NIST Nominal values in bold C pH 1 values added ...
Страница 72: ...72 pH Buffer Tables 06 HACH Nominal values 4 01 7 00 10 01 0 02 at 25 C C pH ...
Страница 73: ...73 pH Buffer Tables 07 WTW Technical Buffers Nominal values in bold C pH ...
Страница 74: ...74 pH Buffer Tables 08 Hamilton Nominal values in bold C pH ...
Страница 75: ...75 pH Buffer Tables 09 Reagecon Nominal values in bold C pH 1 values added ...
Страница 76: ...76 pH Buffer Tables 10 DIN 19267 Nominal values in bold C pH 1 extrapolated ...