Commissioning
Connecting the Cables for Output and Power Supply
The connectors
(A
and
B)
can be screwed off for installation. Connect the
cables according to the overview of terminal assignments and connection.
Solid cables or flexible (stranded) cables with ferrules can be pushed in
directly (push-in terminals). To insert a stranded cable without ferrule, push
down the orange locking tab using a flat-bladed screwdriver. After having
connected the cables, tighten the two screws of the connector.
A
Conductor Cross-Sections
Conductor Cross-Section Min. Max. Unit
Input
1.5
16
mm
2
Single cables with
M5 ring cable lug
Outputs, power supply
0.25 2.5
mm
2
Single cables, stranded or solid,
stranded with ferrule
(with or without collar)
Note:
• The stripping length or length of the cable ferrule (without collar)
should be 10 mm.
• Maximum diameter of the jacket or ferrule collar: 4 mm
• For other diameters: Check if the cables are suitable!