Diaphragm Pump NMS 030 

 

Assembly and function 

KNF Micro BA306116-306115 NMS030-EN 0221 

Operating and Installation Instructions 

5.  Assembly and function 

Only install the pump under the operating parameters and condi-
tions described in Chapter 4, Technical data. 

Observe the safety precautions (see Chapter 3). 

5.1.  Installation 

  Before installation, store the pump at the installation location to 

bring it up to ambient  

Make sure that the installation location is dry and the pump is pro-
tected against water in the form of rain, spray, splashes and drips. 

Protect the pump against dust. 

Protect the pump against vibration and impact. 

Generally speaking, the pump can be installed in any position. 
For maximum accuracy and rapid venting install the pump, at high-
est point in the system to prevent condensate from collection in the 
pump head.

 

 

 

WARNING 

 Risk of physical injury and damage to the pump due 

to automatic start 

  If the pump overheats and the thermal switch / 

electronics stops pump operation, the pumps will 
restart automatically as soon as the motor has 
had time to cool down. 

Take steps to ensure that this cannot produce a haz-
ardous situation 

5.2.  Electrical connection 

Only have the pump connected by an authorized specialist 

Only have the pump connected when the power supply is discon-
nected. 

When connecting the device to a power source, the relevant 
norms, directives, regulations and technical standards must be ob-
served 

Connecting the pump 

  Make sure that the power supply data match the data on the 

motor's type plate. The current consumption can be found on 
the type plate. 

  Connect the motor cables. For electrical data see Chapter 4. 

 

Note the proper polarity. 

For DC motors: 
red motor cable: + 
black motor cable: - 

In the case of brushless DC motors:  
incorrect polarity may damage to the electronics. 

 

Installation location 

Installation position 

Contact and foreign body pro-

tection 

Содержание NMS 030 KPDC

Страница 1: ...observe at all times the relevant instructions to avoid dangerous situations The manual was produced for the serial pumps stated above With customer specified projects pump types starting with PJM or...

Страница 2: ...ential danger It also indicates the possible consequences of fail ure to observe the warning The signal word e g Warning indicates the level of danger Here you will see actions for avoiding the dan ge...

Страница 3: ...he pump head pump housing diaphragm and valves with the me dium The temperature of the medium must lie within the permissible temperature range see Chapter 4 The transferred gas should not contain par...

Страница 4: ...Always ensure adherence to all information stickers on the pumps such as flow direction arrows and type plates and keep stickers in legible condition All replacement parts should be properly stored an...

Страница 5: ...y requirements regarding electro magnetic compatibility in Directive 2014 30 EU if the pump is pow ered by an industrially produced power supply which is used within the specified limits Supply voltag...

Страница 6: ...at atm Pressure l min 2 3 11 Max operating permissible pressure bar g 0 2 Max vacuum mbar abs 200 Tab 5 The pump type NMS030K_DCB brushless stands for Motor voltage 12V 24V 12V 4 24V 4 Supply voltage...

Страница 7: ...put Vctrl Control voltage range DC 0 7 VDC Speed range 300 2800 1 min Resistant to surge voltages up to 28 VDC Input resistance Vctrl 5 3VDC 62 k Threshold voltage 0 11 VDC Speed voltage constants 460...

Страница 8: ...Risk of physical injury and damage to the pump due to automatic start If the pump overheats and the thermal switch electronics stops pump operation the pumps will restart automatically as soon as the...

Страница 9: ...designed to handle the pneumatic data of the pump see Chapter 4 Tech nical data Only use hoses that are suitable for the maximum operating pressure of the pump see Chapter 4 Only use hoses that are c...

Страница 10: ...angers thereof can be prevented by placing a bypass line with a pressure relief valve between the pressure and suction side of the pump Please contact your local KNF partner for further information ww...

Страница 11: ...hort cycles in your application please contact a KNF pump specialist for further information Telephone number see first page Turning the pump on In order to guarantee that the pump can start every tim...

Страница 12: ...ng interval Pump Regular inspection for external damage or leaks Tab 9 7 2 Cleaning 7 2 1 Cleaning the pump The pump if necessary clean outside with a dry wipe Do not use cleaning solvents as these ma...

Страница 13: ...the suction and pressure sides such as hoses valves or filters are causing too much resistance Modify installation check the cross sections of components Hose connections are not leak tight Secure tra...

Страница 14: ...ump see the Chapter Cleaning the pump 4 Send the pump together with completed Health and Safety Clear ance and Decontamination Form to KNF stating the nature of the transferred medium 5 Package the de...

Страница 15: ...Diaphragm Pump NMS 030 Returns KNF Micro BA306116 306115 NMS030 EN 0221 Operating and Installation Instructions 15 KNF worldwide Find our local KNF partner on www knf com...

Страница 16: ...Diaphragm Pump NMS 030 Revisionsbeschreibung KNF Micro BA306116 306115 NMS030 EN 0221 Original Betriebs und Montageanleitung...

Отзывы: