Diaphragm liquid pump NF 1.25
Use
KNF Flodos AG
BA_NF1.25_EN_03_163242
Translation of Original Operating and Installation Instructions, English
3
2. Use
2.1. Intended use
The pumps are intended for transferring and metering liquids.
Owner's responsibility
Only install and operate the pumps under the operating parameters
and conditions described in Chapter 4, Technical data.
Only complete pumps may be taken into service.
Before transferring or metering a medium, check whether the
medium can be transferred danger-free in the specific application
case.
Before using a medium, check the compatibility of the materials of
the pump head, pump housing, diaphragm and valves with the
medium.
The temperature of the medium must lie within the permissible
temperature range (see Chapter 4).
The pumped medium should not contain particles as these can
prevent the pump from working correctly. If this cannot be ensured,
a < 50
m filter with a sufficiently large filter area must be used
upstream of the pump.
2.2. Improper use
The pumps must not be operated in an explosive atmosphere.
For special modifications outside the standard technical specifica-
tions, please contact your KNF technical adviser.
Operating parameters and
conditions
Requirements for