Safety
Chemically-resistant Laboratory pump N820.18
EX
18
Translation of Original Operating Instruction, English, KNF
317428-317425 09/20
Working with
combustible
media and
explosive
atmospheres
Note that the pumps are only suitable for transporting explo-
sive atmospheres and must not be set up in potentially explo-
sive areas.
Make certain that the temperature of the medium is always
sufficiently below the ignition temperature of the medium so
as to prevent ignition or explosion. This also applies for ab-
normal operating situations.
It is important to note here that the temperature of the
medium increases when the pump compresses the medium.
You should therefore ensure that the temperature of the
medium also remains far enough below the ignition tempera-
ture of the medium, even when it is compressed to the maxi-
mum permissible operating pressure of the pump. The maxi-
mum permissible operating pressure of the pump is stated in
the technical data (
).
Where applicable, also take into account external energy
sources (such as radiation sources) that could additionally
heat the medium.
In case of doubt, contact KNF Customer Service.
Environmental
protection
Store and dispose of the pump, as well as all replacement
parts, in accordance with applicable environmental regula-
tions . Observe both the respective national and international
regulations here. This applies in particular to parts that are
contaminated with toxic substances.
Dispose of packaging material that is no longer needed in an
environmentally responsible manner. The packaging materi-
als are recyclable.
Dispose of old devices in an environmentally responsible
manner. Dispose of old devices via suitable collection sys-
tems. Old devices contain valuable recyclable materials.
EU/EC
directives/
standards
The part of the pump that comes into contact with the media
compiles with Directive 2014/34/EU (ATEX).
The pumps comply with directives:
§
2011/65/EU
§
2014/30/EU (EMC)
§
2006/42/EC.
Содержание LABOPORT N820.3FT.29.18G
Страница 55: ...KNF weltweit Unsere lokalen KNF Partner finden Sie unter www knf com DE EN FR ES IT...
Страница 109: ...KNF worldwide You can find our local KNF partners at www knf com...
Страница 165: ...KNF dans le monde entier Vous trouverez nos partenaires KNF ici www knf com...
Страница 222: ...KNF en el mundo Encontrar a nuestros socios KNF locales en www knf com DE EN FR ES IT...
Страница 279: ...KNF nel mondo L elenco dei nostri partner locali KNF disponibile all indirizzo www knf com DE EN FR ES IT...
Страница 333: ...KNF wereldwijd Onze lokale KNF partners vindt u op www knf com...