47
Maintenance and care
Insert new needle seal.
Increasing the pressure stability of
valves
By changing the disk springs inside the valve, the column
switchings can be built up with greater pressure stability.
Steps
Figure
1. Removing the old
needle seal
2. Insert the new nee-
dle seal (B) into the
screw fitting (A).
3. Place the biconical
seal (C) on the new
needle seal and turn
everything into the
syringe connection
at the valve.
4. Screw the screw fit-
ting (A) on the valve
moderately tight.
5. Insert the injection
syringe all the way
into the new needle
seal (C). Loosen the
screw fitting until
the injection syringe
can still be pulled
easily but with
noticeable friction,
straight out.
Fig. 45
Biconical seal, needle seal,
and screw fitting of the
syringe connection
Fig. 46
Schematic display of the
syringe connection
C
B
A
Needle seal
DYNASEAL
bushing
Biconical
sealing ring
Valve
Order number
Maximum pressure
13P/1C
A1378
25 bar
17P/1C
A1379
50 bar
17P/1C with 1/16”
A1379-1
100 bar