Installation
Screw the valve onto the valve drive
Caution!
Risk of mix-up! Always mount the switch or injection
valve onto the matching valve drive. An operation of
the valve with the wrong valve drive can block the
eluent flow. Damage to the device or the system
might be the consequence. Observe the operating
modes!
Fig. 23
Mounting of the valve on the valve drive
1. Check the positions of the DIP switches on the bottom of the
device to make sure that the electrical control of the valve
drive has been set correctly.
2. Switch on the valve drive. The initialization of the valve drive
causes the markings on the drive axle to be on top.
3. Port 1 of the injection valve has to point up! Plug the injec-
tion valve on the drive axle on the valve drive.
4. Screw the injection valve to the adapter plate with Allen
screws M3.
Legend
A. Valve drive
B. Markings for the
true-sided mount-
ing on the drive axle
C. Injection valve
D. Position 1 top to the
true-sided mount-
ing on the drive axle
E. Allen screws (M3)