40
Operating the valves and valve drives
Command lines for terminal program
Binary control
The specification of nominal positions is required if a valve
should be operated without software. A relay is needed, for
example, via a device with different interfaces (interface box).
Connecting the flat ribbon cable with the
connector strip
To enable signal transmission from external devices to the valve,
flat ribbon cables are connected with a connector strip and con-
nected to the
Remote
terminal strip on the device rear of the
valve.
Command
Action
Comment
1, 2, 3, … 6
Valve sets the entered
position
Only 360°
valves
1, 2, … 15, 16,
01, …
Valve sets the entered
position
Only 360°
valves
H, h
Valve sets the position 1
Only 360°
valves; H =
home position
L, l
Valve sets the position
Loading
Only injection
valves
I, i
Valve sets the position
Injecting
Only injection
valves
V, v
Display version of the
program
-
P, p
Display current position
of the valve
-
T, t
Display model of the
valve (type)
-