10
Pump head repair
Removing the pump head
Procedure
1. Remove tubing from the connections of the piston backflus-
hing (A).
2. Remove all connections from the solvent bottles.
3. Unscrew the eluent line (D).
4. Unscrew the outlet fitting 2 (E) and the inlet fitting pressure
sensor (F) so as to remove the capillary.
5. Loosen the opposite pairs of fastening screws (B) on the pump
head evenly and alternately.
6. Hold the pump head by hand, and consecutively pull out all
fastening screws.
7. Remove the pump head.
.
Fig. 1
Pump head
Legend
A Fastening screw
B
Inlet screw fitting
C
Connection for eluent
line
D
Outlet screw fitting
E
Inlet fitting of the pres-
sure sensor
B
B
E
D
C
A