5
Maintenance and Care
35
Next steps
Mounting a pump head.
Attaching Capillaries to the Pump Head
Pay attention to the torques in the chapter Installation (see page 35).
Prerequisite
The pump head has been removed.
Tool
Open-end wrench, sizes 10, 17
Device defect
Damage to the pump head caused by overtightened capillary fittings. Note
the torque of the fittings.
Use 5 Nm torque for stainless-steel fittings.
Use 1 Nm torque for PEEK fittings.
Note:
PEEK fittings can withstand pressures up to 400 bar for 1/16" and
200 bar for 1/8".
Check Valves
Dirty check valves do not open and close properly. They cause pressure fluc-
tuations and irregular flow. If it is impossible to clean the check valves,
replace the whole unit.
Removing the Check Valves
Prerequisite
The pump head has been purged.
The capillaries have been removed.
The pump head has been dismounted.
Procedure
Process
Figure
1. Slide the fitting
1
onto the cap-
illary
2
.
2. Slide the clamping ring
3
onto
the capillary, so the capillary
comes out.
Fig. 31
Screw fitting
3. Using an open-end wrench,
hold the inlet fitting
5
in place.
4. Tighten the fitting
4
at the
pump head.
5. If the capillary cannot be
screwed tight, then use a new
clamping ring!
Fig. 32
Pump head
1
2 3
4
5
Torque
Pump head stainless steel
Torque for inlet and outlet fitting
100 ml
15 Nm
250 ml
15 Nm
500 ml
12 Nm
1000 ml
12 Nm