P 2.1L user manual V6840, version
3.
5
16
Installation
Column defect
Damage to the column due to de-aeration.
Open the venting screw.
Remove the column.
Note:
Before the pump can be used, it must be bleeded.
Result
No air bubbles in the pump head and in the capillaries.
Leak Management
The leak management consists of the leak sensor and the drainage system
(funnels, hoses, nozzles). The drainage system ensures that escaping liquids
flow into a waste bottle. When leaks are registered by the leak sensor, the
LED flashes red. Both the device and the data acquisition via chromatogra-
phy software are stopped.
Prerequisite
The front cover has been removed.
Procedure
Process
Figure
1. Open the venting screw
1
of
the pressure sensor.
2. With the syringe, extract fluid
through the bleed port
2
.
3. If the extracted fluid flows con-
tinuously, stop suction and close
the venting screw.
Fig. 19
Venting screw of the pressure
sensor
1
2
Process
Process
Figure
1. Carefully push the funnel
1
into the center opening of the
capillary guide
2
.
Fig. 20
Funnel and capillary guide
2. Push the long ending of the first
nozzle
4
into the hose
3
.
Fig. 21
Hose and nozzle
1
2
3
4