P 2.1L user manual V6840, version
3.
5
Troubleshooting
29
System Messages
If other system messages are displayed besides those listed below, please
turn the device off and then on. Inform the Technical Support of the manu-
facturer in case the system message repeats itself.
The system messages are in alphabetical order:
When purging, the
pump switches off.
Check if the venting screw on the pressure sensor
is turned up.
Pump does not
transport solvent
Purge the pump head to remove the air bub-
bles.
Clean the ball valves.
Exchange the ball valves.
If the pump head seals are defective, solvent
enters the piston backflushing; inform the
technical support of the manufacturer.
Exchange the pump head
Pressure and flow
rate variations
Purge the pump head to remove the air bub-
bles.
Let the pump run for 1 h.
Always tighten the inlet screw 1 and outlet
screw 1 on the pump head with a torque
wrench and 15 Nm.
Clean the ball valves.
Exchange the ball valves.
Exchange the pump head
Inform the Technical Support of the manufac-
turer.
Pump head leaks
Inspect the inlet and outlet screw fittings of the
pump head.
If the seals are defective, eluent enters the pis-
ton backflushing; inform the Technical support
of the manufacturer.
Exchange the pump head
Flow rate is not cor-
rect
Check the following options:
Check the data for the solvent compressibility
Cleaning the Check Valves
Exchange the ball valves
System message
Solution
A
Auto pump head
type: head data
uninitialized!
Switch the device off and on
Check whether a pump head with RFID recog-
nition has been installed
Repeat the automatic configuration step in the
chromatography software
Remove pump head, clean it and install it again
Содержание Azura V6840
Страница 1: ...Pump P 2 1L User Manual V6840...