P 2.1L user manual V6840, version
3.
5
24
Installation
avoid interference. Therefore, the port number for each device must be
changed and this same number must be entered into the device configura-
tion of the chromatography software. We recommend to use the same port
number for all devices in the same system.
Note:
The port is set to 10001 at the factory. You must use the same num-
bers in the device configuration of the chromatography software as in the
device, otherwise the connection fails.
Process
1. Find out port number and change it on the device.
2. Enter the port number in the chromatography software.
Result
The connection is established.
Setting a Static IP Address
Note:
Please inform yourself about the IT safety standards valid for your lab-
oratory, before changing the LAN settings.
The device is factory set to a dynamic IP address (DHCP). To ensure a per-
manent LAN connection between the chromatography software and the
device, we recommend to set a static IP address for certain applications.
Prerequisites
The device has been switched on.
Mobile Control has been installed and started.
The connection between the Mobile Control and the device has been
established.
Note:
You find further information on LAN settings in the chapter Device
Settings of the Mobile Control User Manual.
Procedure
1. In the Mobile Control, choose
Settings
.
2. On the
General
tab, choose the device name.
3. Under
Network Settings
, choose the setting
Static
1
.
4. Enter the IP address into the text box
IP Address
2
.
5. If necessary, change the subnet mask and the gateway
3
.
6. Click
in the top right corner.
7. Restart the device.
Next steps
In case necessary, go back to the original setting with the [
Reset
] button that
can be found under
Settings
>
General
>
Network Settings
>
LAN Settings
.
Analog Control
Note:
A Control Unit or Mobile Control is required, to change the default
settings and to select the option
ANALOG
in the
SETUP
menu.
Analog ports serve for exchanging analog control signals. Reference point
for the signals is the connector GND.
OUT: Device sends signal.
IN: Device receives signal.
1
2
3