91
Troubleshooting
AZURA® Detector ECD 2.1 Instructions, V6685
Noise: If a significant drop of the noise only is observed it could be sig-
nature of for example pump problem (check valves, air in pump head,
compressibility issues or leaking seals).
Fig. 78
Significant drop in noise
In case no significant drop in noise or cell current is observed service
or replace the cell. In case you still cannot solve the problem contact
your local representative.
Please bear in mind that analytical problems may also be caused by ex-
ternal Influences like temperature, unstable samples. Make sure that the
application was running trouble-free before and that no changes have
been made to the system. A number of causes and possible solutions for
analytical problems is listed below. Contact your local representative if
you need further help.
No detector response
Possible cause
Remedy
No power
Check line voltage setting, plug in power
cord
Power switch off
Turn this switch ON (at the rear panel)
Faulty fuse
Replace fuse
Divergent mains
voltage
Check line voltage
Cell disconnected, or
switched off
Check connection
Output disconnected
Check connection
Fouled WE
Clean WE
Содержание Azura ECD 2.1
Страница 1: ...HPLC Detector ECD 2 1 Instructions Document No V6685 ...
Страница 31: ...27 ECD 2 1 controller AZURA Detector ECD 2 1 Instructions V6685 5 2 Overview of ECD 2 1 screens DC mode ...
Страница 32: ...28 ECD 2 1 controller AZURA Detector ECD 2 1 Instructions V6685 Pulse mode ...
Страница 33: ...29 ECD 2 1 controller AZURA Detector ECD 2 1 Instructions V6685 Scan mode CONFIC menu ...
Страница 34: ...30 ECD 2 1 controller AZURA Detector ECD 2 1 Instructions V6685 DIAG menu ...