64
AZURA® Detector ECD 2.1 Instructions, V6685
10. Technical data
10.1 Environmental, dimensions, weight and
power requirements
Working temperature
10 - 40 °C (indoor use only)
Storage temperature
–25 - +50 °C
Humidity
20 - 80% RH
Safety and EMC
According to EC-directives; Emission Group I
Class A; cMETus approved
Equipment class
1
Installation category
II
Pollution degree
2
Dimensions
43 (D) x 22 (W) x 44 (H) cm = 16.9” (D) x 8.7”
(W) x 17.3” (H)
Weight
max 14.4 kg (32 lbs) without flow cell and
column
Installation
Install upright on flat & smooth surface, keep
space under the detector free (risk of blo-
cking power supply fan).
Power requirements
100-240 VAC, 50/60 Hz, 260 VA, auto-sensing
Mains fuse
2.5 AT / 250V, 5x20 mm, IEC 60127-2
For safety reasons, any other internal fuse
or circuit breaker is not operator accessible,
and should be replaced only by authorized
personnel by the manufacturer. Only use
manufacturer-supplied fuses.
For optimum analytical performance it is recommended that the
ambient temperature of the laboratory be between 20-25 °C and be
held constant to within ± 2 °C throughout the entire working day.
Note: that for optimal performance of the oven the oven temperatu-
re should be set at least 7 degrees higher as ambient temperature.
Содержание Azura ECD 2.1
Страница 1: ...HPLC Detector ECD 2 1 Instructions Document No V6685 ...
Страница 31: ...27 ECD 2 1 controller AZURA Detector ECD 2 1 Instructions V6685 5 2 Overview of ECD 2 1 screens DC mode ...
Страница 32: ...28 ECD 2 1 controller AZURA Detector ECD 2 1 Instructions V6685 Pulse mode ...
Страница 33: ...29 ECD 2 1 controller AZURA Detector ECD 2 1 Instructions V6685 Scan mode CONFIC menu ...
Страница 34: ...30 ECD 2 1 controller AZURA Detector ECD 2 1 Instructions V6685 DIAG menu ...