(1) Disabled: to disable the DO1 output;
(2) High Flow Alarm: DO1 outputs as a high flow alarm when the flow percentage exceeds
the Hi Alm Limit;
(3) EmpPipe Alarm: When the pipe is detected as empty, an alarm signal is output from DO1;
(4) Flow Direction: the DO1 outputs as a flow direction indicator;
(5) Pulse Output: the DO1 outputs pulse signal.
6.1.3.2.19
Hi Alm Limit
High alarm limit value is set in percentage of the upper range of flow rate. The parameter
ranges from 0% to 199.9%. The meter outputs alarm signal when the flow percentage is higher than
this value.
6.1.3.2.20
DO2 Output
User can program the DO2 output by selecting the following options:
(1) Disabled: to disable the DO2 output;
(2) Low Flow Alarm: DO2 outputs as a low flow alarm when the flow percentage is lower
than the Lo Alm Limit;
6.1.3.2.21
Lo Alm Limit
Low alarm limit value is set in percentage of the upper range of flow rate. The parameter
ranges from 0% to 199.9%. The meter outputs alarm signal when the flow percentage is lower than
this value.
6.1.3.2.22
Sensor S/N
Sensor serial number records the information of the sensor equipped with the converter and
ensure them match up when installing.
6.1.3.2.23
Sensor Fact.
The sensor factor is set according to the calibration sheet supplied by the manufacturer.
Usually this factor has been set up by the manufacturer before shipping. It is an important value
that determines the accuracy of measurement. Do not change it without calibration.
6.1.3.2.24 Field Mode
The converter offers three field exciting modes based on the exciting frequency. Mode 1 is the
most-commonly used one and suitable for most cases. Mode 2 and 3 are low-frequency exciting
modes and are better for large size meter to measure water. The calibration should be taken under
the same exciting mode as that used for measurement.
6.1.3.2.25 RevMeas.Enbl: Reverse Measurement Enable
If RevMeas.Enbl is set to ON, the converter displays flow and outputs signals when flow
direction is reversed. If OFF, the converter displays no flow and does not output signals when
reversing.