6.1.3.2.26 Multiplying
This item is a multiplying factor selectable from 0.0000 to 3.9999. When calculating the flow
rate and total, this factor is taken into account. It is often used to measure the flow in the open
channel. If not applied, set the value to 1.0000.
6.1.3.2.27 F. Total Set and R. Total Set
Presetting of forward and reverse total counter is designed to start counting from the existing
reading when replacing a converter or flowmeter. It provides a continuous total flow read which is
convenient for management.
6.1.3.2.28 Input Contrl
Not available.
6.1.3.2.29 Clr Totalizr
Enter the ‘Totalizer Reset Password’ in this menu item and press ENTER to confirm. The
converter clears the three internal counter and restart counting if password matched.
6.1.3.2.30 Clr Tot. Key
The ‘Totalizer Reset Password’ is changeable in this menu item if Level-3 password is entered.
Remind: keep the new password in a safe place.
6.1.3.2.31 Date –y/m/d and Time-h/m/s
These items are used to change the internal real time clock if equipped.
6.1.3.2.32 Password L1, Password L2 and Password L3
To change the Level-1 to Level-3 passwords, use Level-4 or higher level password to enter and
change these two items.
6.1.3.2.33 Current Zero and Current Max
Adjust the current output zero point and upper range value as detailed in Sec. 2.7. It is not
suggested that user make any adjustment since it has been setup to the best condition by the
manufacturer.
6.1.3.2.34 Meter Factor
This factor is used by the manufacturer to normalize the excitation current and amplifier signal
of the converter. DO NOT change it.
6.1.3.2.35 Convtr S/N
This serial number records the manufacturing date and code of converter. DO NOT change it.
6.1.3.2.36 Sys Reset
This item is reserved for the manufacturer to re-initialize the converter. After system resetting,
all settings are set to default values automatically.