NOTICE
The content of this document is the intellectual property of Kistler-Morse
®
Corporation. Any
reproduction or translation of this document without the written authorization of a Kistler-Morse
®
corporate officer is prohibited.
CAUTION
Follow these rules if welding is done on the vessel after installation of the Microcell™ system. The
electrical current of the welder may pass through the Microcell™, causing damage to the sensor and
possibly to the signal processor. To avoid damage, follow these precautions:
1. Disconnect the Microcell™ cables from the signal processor.
2. Ground the welder as close to the welding joint as possible. The welding ground must be between
the Microcell™ and the weld joint to prevent the welding current from going through the Microcell™
to earth ground.