MKIII OPERATOR'S MANUAL
64
Approximate run times for the following measurements are shown below:
HG (without
7-9 min.
SL(without B2)
9 min
DS
3 min
ZS
4 min
UV
8 min
UX
6 min
UM
Indeterminate must be aborted with
HOME key or new schedule entry
FM
11 min
Writing Schedules:
•
Determine the Scientific Objectives, and draw up a table of Solar Zenith Angle vs.
Commands.
•
Use the SA command to assist in relating to local time.
•
From the Brewer main menu, send the Command SE.
•
a table will appear on the screen with two columns, one for SZA, and one for the commands
to be executed at the SZA.
•
starting with the first SZA after solar midnight (it will probably be -ve), enter the SZA in the
left hand column of the table, and enter the commands to be executed at that angle in the
right hand column.
•
carry on through the day, remembering that the SZA changes to +VE after noon (no sign
defaults to +ve).
•
when the schedule is finished, press Ctrl+End and follow the instructions for saving the
schedule.
•
the schedule can be tested using the SK or SKC routines.
Содержание Brewer Mk III
Страница 1: ...Operators manual Spectrophotometer Brewer MkIII...
Страница 3: ...ii...
Страница 5: ...2...
Страница 7: ...4...
Страница 25: ...MKIII OPERATOR S MANUAL 22...
Страница 27: ...MKIII OPERATOR S MANUAL 24 Figure 3 2 Brewer Equipment Setup Figure 3 3 Brewer Spectrophotometer Tracker Tripod...
Страница 29: ...MKIII OPERATOR S MANUAL 26 Figure 3 4 Brewer Spectrophotometer Tripod...
Страница 63: ...MKIII OPERATOR S MANUAL 60...
Страница 83: ...MKIII OPERATOR S MANUAL 80...
Страница 91: ...MKIII OPERATOR S MANUAL 88...
Страница 93: ...MKIII OPERATOR S MANUAL 90...
Страница 97: ...MKIII OPERATOR S MANUAL 94...
Страница 129: ...MKIII OPERATOR S MANUAL 126...
Страница 135: ...MKIII OPERATOR S MANUAL 132...