MKIII OPERATOR'S MANUAL
10
Figure 2.2: View of Brewer with Covers Removed
2.1
SPECTROPHOTOMETER
Table 2.1: Spectrophotometer Specifications
UV wavelengths:
‘ozone’ wavelengths (nm): 303.2 (Hg slit) , 306.3, 310.1,
313.5, 316.8, 320.1
Mercury-calibration
302.15 nm
Resolution
0.6 nm in UV
Stability
±0.01 nm (over full temperature range)
Precision
0.006 ±0.002 nm
Measurement range
286.5 nm to 363.0 nm
Exit-slit mask cycling
0.12 sec/slit, 1.6 sec for full cycle
O
3
Measurement accuracy
±1% (for direct-sun total ozone)
Ambient Operating temperature
range (for more information see
appendix F, Thermal Test)
0
°
C to +40
°
C (no heater)
-20C to +40
°
C (with heater option)
-50
°
C to +40
°
C (with complete cold weather kit)
Physical dimensions (external
weatherproof container)
Size: 70 by 46 by 34 cm
Weight: 34 kg (75 lbs)
Power requirements
Brewer and Tracker
90 to 264 VAC
47 to 440 Hz
Содержание Brewer Mk III
Страница 1: ...Operators manual Spectrophotometer Brewer MkIII...
Страница 3: ...ii...
Страница 5: ...2...
Страница 7: ...4...
Страница 25: ...MKIII OPERATOR S MANUAL 22...
Страница 27: ...MKIII OPERATOR S MANUAL 24 Figure 3 2 Brewer Equipment Setup Figure 3 3 Brewer Spectrophotometer Tracker Tripod...
Страница 29: ...MKIII OPERATOR S MANUAL 26 Figure 3 4 Brewer Spectrophotometer Tripod...
Страница 63: ...MKIII OPERATOR S MANUAL 60...
Страница 83: ...MKIII OPERATOR S MANUAL 80...
Страница 91: ...MKIII OPERATOR S MANUAL 88...
Страница 93: ...MKIII OPERATOR S MANUAL 90...
Страница 97: ...MKIII OPERATOR S MANUAL 94...
Страница 129: ...MKIII OPERATOR S MANUAL 126...
Страница 135: ...MKIII OPERATOR S MANUAL 132...