CALIBRATION AND MAINTENANCE
KI7000 UM-5
-
38 -
There are no internal user adjustments. Calibration is performed
without opening the instrument.
All Calibrations
To enable calibration mode, open the battery compartment, remove
the anti-tamper label, and insert a 2.54 mm (0.1”) pitch programming
shunt. Manipulation of the shunt is easier with needle nose pliers.
The instrument will display ‘CAL’ and installed options.
View Inside Battery Compartment
Known calibration constants can be re-entered directly without using
other equipment. This is useful in case old calibration constants are
to be put back.
Before commencing calibration:
•
Clean all connectors very carefully.
•
Ensure that all devices have been at a stable room
temperature for over an hour, and that the light source is fully
warmed up at the wavelength to be calibrated.
•
Ensure that all installed batteries are in good working
condition.
•
When calibration is complete, remove the calibration shunt,
and place an anti-tamper label over the aperture in the
battery compartment.
•
Do no forget to update your calibration records, and schedule
the next calibration service.
System Rollback
(version 5.0 and above)
In the event of wrong calibration constants being entered and stored,
or the instrument becoming otherwise corrupted, all factory settings
can be restored. To restore all factory settings, first enter ‘CAL’ mode
then press and hold [MR] for 3 seconds.
Содержание KI7300 Series
Страница 1: ......
Страница 57: ...QUICK REFERENCE GUIDE KI7600C Series Optical Power Meter KI7000 UM 5 55 ...
Страница 59: ...QUICK REFERENCE GUIDE KI7400C 7800C Series Optical Light Source KI7000 UM 5 57 ...
Страница 62: ...QUICK REFERENCE GUIDE KI7300C 7700C Series Loss Test Set KI7000 UM 5 60 ...
Страница 64: ...NOTES KI7000 UM 5 62 ...
Страница 65: ...NOTES KI7000 UM 5 63 ...
Страница 66: ...NOTES KI7000 UM 5 64 ...