![Kingfisher KI 7340 Series Скачать руководство пользователя страница 59](http://html1.mh-extra.com/html/kingfisher/ki-7340-series/ki-7340-series_operation-and-maintenance-manual_1970295059.webp)
KI 734x/774x UM-12
57
PERFORMANCE VERIFICATION TESTS
Model:_______________________________
Report No.:______________ Date:________________
Total Uncertainty / Accuracy Test
Test Wavelength = nm
Setting
Number
Power Meter
reference value
Attenuator
Setting
Minimum specification
(-0.3 dB of Ref.)
DUT
measurement results
Maximum specification
(+0.3 dB of Ref.)
1.
(~-10.00 dBm)
dB
(~-10.30 dBm)
dBm
(~-9.70 dBm)
2.
(~-20.00 dBm)
dB
(~-20.30 dBm)
dBm
(~-19.70 dBm)
3.
(~-30.00 dBm)
dB
(~-30.30 dBm)
dBm
(~-29.70 dBm)
4.
(~-40.00 dBm)
dB
(~-40.30 dBm)
dBm
(~-39.70 dBm)
5.
(~-50.00 dBm)
dB
(~-50.30 dBm)
dBm
(~-49.70 dBm)
Measurement Uncertainty
dB
Table 4: Accuracy Test Record for KI7340C/7740C Series, Power Meter section
Note 1
: This is for the KI7343C-InGaAs. For the KI7344C- H series instruments, increase by ±0.2 dB.
For the KI 7600A-H3, increase by ± 0.1 dB.