![Kingfisher KI 7340 Series Скачать руководство пользователя страница 53](http://html1.mh-extra.com/html/kingfisher/ki-7340-series/ki-7340-series_operation-and-maintenance-manual_1970295053.webp)
KI 734x/774x UM-12
51
PERFORMANCE VERIFICATION TESTS
Performance Verification of One Way Autotest function
One Way Range and Linearity Test
Test described below allows checking One Way Autotest range and
linearity for one loss test set and also verifies tone detector operation.
To test power meter absolute accuracy and linearity, refer to Accuracy
Test, Power Meter Section.
1.
Connect the equipment as shown in Figure 11. A single LTS can
also be used, with connections to both ports.
LTS Attenuator LTS
Figure 11. Test set up for One Way Range and Linearity Test
2. Switch on all instruments.
3. Set the Light Source to 1310nm and allow to warm up. Press
[Autotest].
4. On the attenuator, set the wavelength to1310nm and attenuation
to 5 dB.
5. Make sure that Autotest is updating. Note the measured optical
power at 1310 nm.
6. For singlemode instruments, repeat the power readings for
attenuator settings of 15 dB, 25 dB and 35 dB. For instruments
with H series meter, repeat the readings for an attenuator setting
of 15 dB only.
7. For multimode instruments, repeat the power readings for
attenuator settings of 15 dB, 20 dB.
8. Repeat the above for 1550nm as required, however, this test
typically only needs performing at one wavelength.
Содержание KI 7340 Series
Страница 1: ...OPERATION MAINTENANCE GUIDE KI 7340 Series Two Way Test Set ORL KI 7740 Series Two Way Loss Test Set ...
Страница 66: ...KI 734x 774x UM 12 64 KI7340C 7740C SERIES TWO WAY LOSS TEST SET ...
Страница 68: ...KI 734x 774x UM 12 66 NOTES ...
Страница 69: ...KI 734x 774x UM 12 67 NOTES ...
Страница 70: ...KI 734x 774x UM 12 68 NOTES ...