KI 734x/774x UM-12
50
PERFORMANCE VERIFICATION TESTS
Performance Verification of Two Way Autotest function-
continued:
Two Way Range and Linearity Test
1.
Connect the equipment as shown in Figure 10.
LTS Attenuator LTS
Figure 10.Test set up for Two Way Range and Linearity Test
2. Switch on all instruments.
3. Set the Light Source on one of the instruments to 1310nm and
allow to warm up. Press [Autotest].
4. On the attenuator, set the wavelength to1310nm and attenuation
to 5 dB.
5. Make sure that Autotest is updating. Note the measured optical
power at 1310 nm.
6. For single mode instruments, repeat the power readings for
attenuator settings of 15 dB, 25 dB and 35 dB. For instruments
with H series meter, repeat the readings for an attenuator setting
of 15 dB only.
7. For multimode instruments, repeat the power readings for
attenuator settings of 15 dB, 20 dB.
8. Repeat the above for 1550nm as required, however, this test
typically only needs performing at one wavelength.