KI 734x/774x UM-12
46
PERFORMANCE VERIFICATION TESTS
Power Meter section
Accuracy Test
1.
Connect the equipment as shown in Figure 5:
LTS / Power Meter Attenuator Source
Figure 5. Test set-up for the Power Meter section Accuracy Test
2. Switch on all instruments.
3. Set all instruments to 1310 nm.
4. Change the attenuation of the attenuator until the reference power
meter displays -10.00 dBm. Note the attenuator setting in setting
1 of Table 4.
If the light source is not powerful enough to give -4.00 dBm, set
the attenuator to 2.5 dB and correct the appropriate values in the
test report.
Repeat the above for reference power meter readings of -20dBm,
-30 dBm, -40 dBm and -50 dBm (last reading is not relevant to H
series power meters).
5. Measure the DUT:
Re-connect the attenuator output cable to the DUT power meter
port, and select the 1310nm on DUT.
Set the attenuator to its value for setting 1. Note the displayed
power level of the DUT in the test record.
Repeat the above for attenuator settings 2~5.
6. Repeat the Power Meter Accuracy Test at 1550 nm.
Light Source section
Output Power (CW) Test
1.
Connect the equipment as shown in Figure 6.
Power Meter LTS / Source
Figure 6. Test set-up for Output Power (CW) and Short Term Stability Tests
Содержание KI 7340 Series
Страница 1: ...OPERATION MAINTENANCE GUIDE KI 7340 Series Two Way Test Set ORL KI 7740 Series Two Way Loss Test Set ...
Страница 66: ...KI 734x 774x UM 12 64 KI7340C 7740C SERIES TWO WAY LOSS TEST SET ...
Страница 68: ...KI 734x 774x UM 12 66 NOTES ...
Страница 69: ...KI 734x 774x UM 12 67 NOTES ...
Страница 70: ...KI 734x 774x UM 12 68 NOTES ...