ORIGINAL INSTRUCTIONS
Model:
Dispenser Rx
Serial Number:
5780/14
Instructions Manual Dispenser Rx Tablet and Capsule Counting Machine Edition v2.1
30
GENERAL MAINTENANCE AND CLEANING (Continued.)
Cleaning
Clean stainless steel and plastic components using a cloth damped in hot water and finish off
wiping with a clean dry cloth.
Remove grease or oil from metal parts with methylethyketone (MEK) or carbon tetrachloride (CTC)
used sparingly and observing safety precautions associated with these solvents.
Recommended lubricants for general use
Grease – Lithium based grease (JIS No.2), Urea based grease (JIS No.2) or Albania Grease No.2
(Showa Shell Sekiyu)
Oil – Sliding surface oil or turbine oil ISOVG32 ~ 68, Tonner Oil (Showa Shell Sekiyu) or Vactra
No.2S (Mobile Oil)
CAUTION
Failure to observe the instructions could result in damage to the
equipment, associated equipment and processes.
WARNING
Mixing different types of grease is dangerous in some cases.