ORIGINAL INSTRUCTIONS
Model:
Dispenser Rx
Serial Number:
5780/14
Instructions Manual Dispenser Rx Tablet and Capsule Counting Machine Edition v2.1
20
OPERATION: Advanced Settings Screen
Units of Time
•
s = second. 1s = 1 second
•
ms = millisecond. 1ms = 0.001s
•
us = microsecond. 1us = 0.000001s
•
The
ADVANCED SETTINGS SCREEN
controls additional functions which directly effects
the operation of the Count Sensor and Changeover Flap.
•
Adjusting the
PRODUCT LENGTH
changes the amount of time the Count Sensor expects
to be blocked by a single piece of product as it falls past. Adjusting this incorrectly can
cause over or under counting.
•
The
PRODUCT GAP
Determines the distance between the current falling and following
Tablet or Capsule. Having a gap which is too short may cause an over count, if a gap is too
long it will cause an under count.
•
The
FLAP DELAY
becomes active after the final tablet or capsule in the target count falls
past the Count Sensor. Changing the
FLAP DELAY
will increase or decrease the amount
of time a that must elapse after the final tablet or capsule has fallen past the Count Sensor
for the Flap to change its position.
•
The
FLAP DURATION
parameter controls the speed of the flap movement.