Housing

..................

ABS shock-proof
IP54

Keypad

...................

Metal-coated, 5 keypads, 1 joystick 

Conformity

.............

Electromagnetic compatibility
(as per NF EN 61326-1)

Power supply

.........

4 alcalines batteries 1,5V LR6

Ambient

...................

Neutral gas

Operating temp.

....

from 0 to +50°C

Storage temp.

........

from -20 to +80°C

Auto shut-off

..........adjustable

 from 0 to 120 min

Weight

......................

340g

Languages

.............

French, english

4

Technical features

Specifications

I – Technical specifications

Pressure module

MP200 Connection (See p.6)

Display

Piezoresistive sensor

Overpressure allowed ±500 Pa : 250 mbar

Overpressure allowed ±2,500 Pa : 500 mbar

Overpressure allowed ±10.000 Pa : 1,200 mbar

Overpressure allowed ±500 mBar : 2 bar

Overpressure allowed ±2,000 mBar : 6 bar

Measuring units

Measuring range

Accuracy*

Resolutions

Graphic display 128x128 pixels

Dim. 50 x 54 mm, blue blacklit,

Display of  6 measurements (including 4 simultaneously)

Pa, mmH

2

O, In WG, 

mbar, hPa, mmHg,

 DaPa, kPa, bar, PSI

V, mA

From 0 to 2.5 V
From 0 to 10 V
From 0 to 4/20 mA 

m/s, fpm, Km/h, mph

From 2 to 5 m/s
From 5.1 to 100 m/s

±0.3 m/s 
±0.5% of reading ±0.2m/s

0.1 m/s
0.1 m/s

From 0 to 99,999m

3

/h

±0.2% of reading ±1% PE

1 m

3

/h

m

3

/h, cfm, l/s, m

3

/s

Air velocity

CURRENT / VOLTAGE

PRESSURE

Airflow

Airflow

Air velocity

PITOT TUBE

DEBIMO BLADES

m/s, fpm, Km/h, mph

Frpm 4 to 20 m/s
From 21 to 100 m/s

0.1 m/s
0.1 m/s

From 0 to 99,999m

3

/h

1 m

3

/h

m

3

/h, cfm, l/s, m

3

/s

0.001 V
0.01 V
0.01 mA

±0.3 m/s
±1% of reading ±0.1m/s

From -200 to 1,300°C
From -100 to 750°C
From -200 to 400°C

±1.1°C or ±0.4% Reading value**

±0.8°C or ±0.4% Reading value**

±0.5°C or ±0.4% Reading value**

0.1 °C
0.1 °C
0.1 °C

°C, °F

K :

J :

T :

+

°C, °F

ppm

From -20 to +80°C 

±0.3% of reading ±0.25°C

0.1 °C

From 0 to 100 ppm
From 100 to 500 ppm

±3 ppm

0.1 ppm

CO / Temperature

CO

Temp.

THERMOCOUPLE  (See related datasheet)

±2mV

±10mV

±0.01mA

±0.2% of reading ±1% PE

±3% of reading 

From 0 to ±500 Pa
From 0 to ±2,500 Pa

±0.2% of reading ±2Pa

1Pa

From 0 to ±10.000 Pa

±0.2% of reading ±10Pa

1Pa

From 0 to ±500 mbar

±0.2% of reading ±0.5mbar

0.1mbar

From 0 to ±2,000 mbar

±0.2% of reading ±2mbar

1mbar

0.1 Pa from -100 to + 100 Pa,  

1 Pa beyond

± 100 Pa : ±0.2%  of reading ±0.8Pa,
beyond ±0.2%  of reading ±1.5Pa,

*All accuracies indicated in this document were stated in laboratory conditions and can be guaranteed for measurements carried out in the same conditions, or carried out with required compensation.
** The accuracy is expressed either by  a deviation in °C  or by a percentage of the value concerned. Only the bigger value is considered.

Gas leak

ppm

%LEL

%VOL

From 0 to 10 000 ppm 
(GPL : 0-1800)

From 0 to 1%VOL

From 0 to 20%LEL

±20% of full scale at 20 °C at 

65 %HR ± 5 %HR

1 ppm

0,001 %VOL

0,01 %LEL

actoolsupply.com

actoolsupply.com

Содержание MP 200 G

Страница 1: ...MP 200 Manometer Supplied with Calibration certificate actoolsupply com actoolsupply com actoolsupply com...

Страница 2: ...lsupply com Kimo MP 200 P Thermo Anemometer Manometer Kimo MP 200 M Thermo Anemometer Manometer Kimo MP 200 G Thermo Anemometer Manometer Kimo MP 200 H Thermo Anemometer Manometer Kimo MP 200 HG Therm...

Страница 3: ...tion 9 Point point average 9 Automatic average 9 Automatic point point average 10 Configuration 10 Thermocouple type 10 Display 10 Units 10 Integration 10 Compensation 10 Pressure system 10 Solenoid v...

Страница 4: ...TUBE DEBIMO BLADES m s fpm Km h mph Frpm 4 to 20 m s From 21 to 100 m s 0 1 m s 0 1 m s From 0 to 99 999m3 h 1 m3 h m3 h cfm l s m3 s 0 001 V 0 01 V 0 01 mA 0 3 m s 1 of reading 0 1m s From 200 to 1 3...

Страница 5: ...w Module connection mini Din C1 connector mini Din C2 connector Graphic display Hour Value Unit Circular menu Memory status used Battery level Keypad On Off key Access keys for circular menu Escape ke...

Страница 6: ...and and thermocouple temperature measurement on Tc1 channel with wire thermocouple probes equipped with a miniature male connector Wire probes with Smart plus system Wire probes with Smart plus system...

Страница 7: ...h access keys or with arrow keys Measurement Communication interrupted Check probes connection Select a sub menu with access keys or with arrow keys Select a sub function with access keys or with arro...

Страница 8: ...velocity To select vent Type press OK or Select Lx W or Diam or K factor with arrow buttons and Confirm with OK If K factor is selected you must enter value You can choose a K factor already registere...

Страница 9: ...max and min values standard deviation average of each channel and e numbers of measuring points will be displayed If you want to see all values select Visu and scroll with and Point point average Pre...

Страница 10: ...in more precise results Click on OK or to enter into the sub function Select or signs with and with and then pass on the first digit with Enter the first digit then move to the next one with Confirm w...

Страница 11: ...dataset An automatic dataset is composed of measuring points with interval of time a Click on OK or to enter sub function b Select Auto with and Confirm wih OK c Select Name with and Confirm wih OK o...

Страница 12: ...enus 4 Delete all datasets Select Delete with and Confirm wih OK Click on OK or to enter into sub function Enter the day with and then move to the next digit with Repeat this operation for the month y...

Страница 13: ...urity code Click on OK or and the code appears Enter the first digit of the code with and then move to the next one with Confirm wih OK Contrast Backlit Key locking Code RF logging This sub function a...

Страница 14: ...nstant level of quality of your measurements In regards of Quality insurance norms we recommend that the instruments are checked once a year KIMO Instruments have 1 year guarantee for any manufacturin...

Страница 15: ...actoolsupply com actoolsupply com...

Страница 16: ...We reserve the right to modify the characteristics of our products without prior notice Once returned to KIMO required waste collection will be assured in the respect of the environment in accordance...

Отзывы: