background image

KPC 3k and 15k

 

MAN08152022

 

Unpacking and setup 

 

Remove the scale from the box and place it on a firm, level surface. 

 

 

Avoid locations with rapid temperature changes, excessive dust, moisture, air currents, vibrations, 
electromagnetic fields, heat or direct sunlight.

 

 

Adjust the four levelling legs by turning them so that the air bubble is centered on the black circle of the 
level bubble. All levelling legs should be touching the surface. 

 

 

 

NOTE: Ensure that the scale is level each time its location is changed.

 

 

Before using the scale for the first time, the internal rechargeable battery should be fully charged for at 
least 12 hours. 

 

 

Connect the supplied AC adapter to the power input receptacle underneath the scale. Plug the AC adapter 
into a power outlet and the battery will begin charging.

 

 

If the scale will be stored or transported in the future, save the packaging material to ensure the best 
possible protection for the scale. 

 

 

Weighing operation

s

 

 

KEYBOARD AND DISPLAY 

 

Front Panel 

 

KEY

 

DESCRIPTION

 

 

Press to turn on / off the scale.

 

The scale will first run a self-test and then enter into weighing mode.

 

MODE

 

Press to enter into the configuration setting menu.

 

When in weighing mode, use to change the weighing units.

 

→T←

 

Press to tare to deduct the container weight; net weight will display. 

 

→0←

 

Press to zero the scale. Range of zero is ±2% FS of full scale.

 

Содержание KPC 15K

Страница 1: ... Owner s Manual KPC 3K AND KPC 15K OPERATION MANUAL E MAN08152022 Ce manuel est disponible en français à http www kilotech com K850137 KPC 3K 3KG X 0 1G 6LB X 0 0002LB GTIN 889605250137 K850138 KPC 15K 15KG X 0 5G 30LB X 0 001LB GTIN 889605250138 ...

Страница 2: ...efects in material and workmanship for a period of one year for the scale and 3 months for the adapter from the date of purchase Please retain your receipt as proof of purchase Warranty expiration on products returned without proof of purchase will be determined from the manufacturing date code or the date of shipment to the authorized dealer Coverage may be invalidated as this date is previous to...

Страница 3: ...ad beyond max cap the maximum weight limit Do not pull the plug by its cord when unplugging Do not use the scale near fire or corrosive gas Do not spill water or immerse in water Avoid placing the scale near a heater It could cause a fluctuation to the indicator reading Always make sure to use a Kilotech approved adapter Do not drop the scale or apply a sudden shock Do not drag the scale Avoid exp...

Страница 4: ...power off function YES Zero tracking zero error indication overload alert function YES Error signal YES Counting Feature YES Power Input AC110V 240V 50HZ Output AC9V 12V 500mA DC 6V4Ah recharging battery Power dissipation 0 1VA Product Dimensions 26cm 10 25 x 20cm 8 x 8cm 3 25 Platter Size 20cm 8 18cm 7 Operating Temperature Humidity 0 40 85 RH Minimum load cell resolution 0 1uV Maximum resolution...

Страница 5: ... battery should be fully charged for at least 12 hours Connect the supplied AC adapter to the power input receptacle underneath the scale Plug the AC adapter into a power outlet and the battery will begin charging If the scale will be stored or transported in the future save the packaging material to ensure the best possible protection for the scale Weighing operations KEYBOARD AND DISPLAY Front P...

Страница 6: ...ress MODE to confirm and enter into next step 02 Press MODE bp On Buzzer On or OFF setting Press 0 to change data Press MODE to confirm and enter into next step 03 Press MODE Pr AB Backlight setting A 0 2 can be selected 0 backlight always off 1 when weighing and the weight is stable backlight will turn off after the numbers of seconds set in the B setting below 2 when zero displayed the backlight...

Страница 7: ... X 1 means after stable status data sent out 05 Press MODE PLo 20 Rechargeable battery use setting Press MODE to skip this step and enter into next step Press 0 to change setting digits 0 40 can be selected Ex 20 means when rechargeable battery indicator is 20 scale will power off automatically Press MODE to confirm and enter into next step 06 Press MODE LCd 5 LCD brightness setting Number 3 15 ca...

Страница 8: ...ng stable lock setting X 0 unlocking X 1 locking after common stable X 2 press 0 to lock and unlock X 3 weight value of buzz locking press 0 to unlock Press 0 key to change data Press MODE to confirm and enter into next step Press MODE FLT1 X Locking time setting only FLTO 1 effective X 0 means after stable locking X 1 9 means after weighing stable 1 9 seconds locking Press 0 key to change data Pr...

Страница 9: ... SP Calibration status setting 02 Press T CAL 0 Enter into calibration status 03 Press MODE 1 Entering auto zero calibration after then indicator show the max capacity ex 15000g 04 Press MODE XXXXXX eg 10000g When Max Capacity weight display ex 15000g Load the test weight 1 3 max Capacity ex 10000g After stable symbol STB display Press T to select digit position and press 0 to set the digits from ...

Страница 10: ...on the pan ex 50 pieces After the icon is displayed press and hold the MODE key for three seconds The display will read 000000 and the PCS icon will be active Press T to select the position and 0 to set the sample size until the display reads 000050 Then press MODE to see n 50 displayed You may need to add more items if you get the Err 11 code since the weight of the item may be too light or out o...

Отзывы: