KPC 3k and 15k
MAN08152022
09
Press MODE
“- Zero -”
Zero weight change tracking.
When in weighing stable status, changing weight< setting
weight, original weight will maintain, false operation.
Press →T←
“E1=0”
Division setting:
E1=0.0d, setting range:0.0d~2.0d can be selected.
press “→0←” key to change data.
*Press “MODE” to confirm and enter into next step.
Press MODE
“E2=0”
Valid time setting.
(0`9 can be selected)
press “→0←” key to change data.
*Press “MODE” to confirm and enter into next step.
10
Press MODE
- PASS -
Press →0←, return back to weight mode (modify done)
11
Press →0←
0.000
Return back to weight mode.
CALIBRATION
Press MODE and “→T←” together for 3 seconds till “= = = = = =”display then release, parameter as follow
will show on the display.
Step
Operation
Screen
Display
Remark
01
CAL SP
Calibration status setting
02
Press →T←
CAL 0
Enter into calibration status.
03
Press MODE
1.
- - - - -
-
Entering auto zero calibration, after then indicator show
the max. capacity ex:15000g.
04
Press MODE
XXXXXX
(eg:10000g)
When Max. Capacity weight display, ex: 15000g. Load
the test weight (≥1/3 max. Capacity), ex: 10000g.
After stable symbol “STB” display, Press →T← to select
digit position and press →0← to set the digits from
0~9.eg:10000g. Press MODE to confirm.
05
Press MODE
- - - - - -
Handling relative weight
10000
The weight should be same as the test weight, if not,
please do the calibration again as per the above steps.
BASIC INFORMATION DISPLAY
Press MODE and “→T←” together for 3 seconds till “= = = = = =”display then release, parameter as follow
will show on the display.
Step
Operation
Screen Display
Remark
01
“CAL SP”
Calibration status setting
02
Press →T←
“-SEt-”
System parameter setting.
03
Press MODE
“-A-D-”
AD data
04
Press →T←
“33511”
Current AD inner number
05
Press →0←
“bU=6.78”
Current battery