PMC-A
73
6
M
a
inte
na
nc
e
■
Ammeter at full scale
11.
Set the o
u
tp
u
t c
u
rrent (a val
u
e calc
u
lated from the
external DVM reading and sh
u
nt resistance) to the
rated c
u
rrent.
12.
Adj
u
st I METER FS so that the ammeter reading on
the power s
u
pply is eq
u
al to the rated c
u
rrent.
Type I: Variable resistor 2
Type II: Variable resistor 5
■
Preset current indicator at full scale (Type I model
only)
13.
Adj
u
st I LIMIT FS so that the ammeter reading on the
power s
u
pply is eq
u
al to the val
u
e calc
u
lated from
the external DVM reading and sh
u
nt resistance when
the LIMIT switch is pressed while delivering rated
c
u
rrent.
Type I: Variable resistor 5
Содержание PMC 110-0.6A