Alarm
The alarm detected by single unit of the PMC-A can be also
detected under the master-slave parallel operation.
Connection and setup procedure
1.
T
u
rn off the OUTPUT and PO
W
ER switches on all
power s
u
pplies that are to be connected in parallel.
2.
Choose the power s
u
pply that will be the master.
3.
Set the OVP (overvoltage protection) trip point on the
master and slave power s
u
pplies.
In parallel operation, set the OVP trip point not only on the mas-
ter power supply but also slave power supplies. However, set the
OVP trip point of the slave power supplies slightly higher than
that of the master power supply, so that the OVP function of the
master power supply is activated first.
4.
Connect and set
u
p each power s
u
pply as shown in Fig.
4-5. Set the S5 switch on all slave power s
u
pplies to
the down (0) position.
As for handling J1 terminal, see page 45, Fig. 4-4 Connection to
J1 terminal.
Fig. 4-5 Master-slave parallel connection and
set
u
p
–
+
1
0
S5
2
3
1
J1
–
–
+
+
1
0
S5
2
3
1
J1
PMC-A master unit
PMC-A sla
v
e unit
To J1-2 of the next sla
v
e unit
Load
Connect the short
b
ar to either the
negati
v
e or positi
v
e output terminal only
on the master unit
Up
Do
w
n