PMC-A
49
4
Bas
ic O
p
er
at
io
n
C.V LED lights on the master power supply panel.
■
End procedure
1.
T
u
rn off the OUTPUT switch on the master power
s
u
pply.
2.
T
u
rn off the OUTPUT switch on the slave power s
u
p-
plies.
3.
T
u
rn off the PO
W
ER switch on the slave power s
u
p-
plies.
4.
T
u
rn off the PO
W
ER switch on the master power
s
u
pply.
When the Output voltage cannot be set. (Type I
model only)
When the output voltage cannot be set with the master unit or a
few voltages are output at 0 V setting, adjust the variable resis-
tor for calibration in the following procedure. See page 67, Test
equipment required for calibrationfor the test equipment
required and environment. See Fig. 6-3 for the connection of
the equipment.
・
Do not set the o
u
tp
u
t c
u
rrent greater than or eq
u
al
to 105 % of the rated c
u
rrent for this adj
u
stment. If
yo
u
do, the power s
u
pply may malf
u
nction
.
■
Adjustment procedure
1.
Set the o
u
tp
u
t c
u
rrent to 0 A.
When operating by the front panel (local control), turn the
CURRENT knob counterclockwise all the way. When using ana-
log remote control, set the control signal to 0 V or 0
Ω
.
2.
T
u
rn on the OUTPUT switch.
CAUTION
Содержание PMC 110-0.6A