background image

10

PAV LAN

LAN configuration

2. Network Connection

Control functions that can be used simultaneously with LAN

On a PAV with LAN, you can use local mode and external control simultaneously with LAN.
However, LAN cannot be used simultaneously with serial communication (USB/RS232/
RS485) remote mode. For the serial communication setup procedure, see the USB/RS232/
RS485 communication interface manual.

Local mode (front panel setup)

Even when connected to LAN, the PAV can be configured using the front panel knobs and
keys.

Switching to local mode

When the PAV is in LAN remote mode, the front panel REM LED is lit. Press REM to turn off
the REM LED and switch to local mode.
If the PAV does not switch to local mode even when you press REM, perform one of the fol-
lowing procedures.

External control

Even when connected to LAN, you can control and monitor the PAV series using external
analog signals through the PAV’s J1 and J3 connectors.

Procedure (switching to local mode)

Description

• Stop sending commands.

Press REM.

If the PAV continuously receives configura-
tion change commands through LAN, the
PAV switches to remote mode each time it
receives a command.

• SYSTem:REMote[:STAte] LOC/0

Send this command.

The PAV may be set to local lockout mode 
through LAN.

• Turn the POWER switch off and then back 

on.
Press REM.

Содержание PAV10-20

Страница 1: ...64 PAV60 7 PAV100 4 600 W type PAV10 60 PAV160 4 PAV20 30 PAV320 2 PAV36 18 PAV650 1 PAV60 10 PAV100 6 800 W type PAV10 72 PAV160 5 PAV20 40 PAV320 2 5 PAV36 24 PAV650 1 25 PAV60 14 PAV100 8 Regulate...

Страница 2: ...ufficient basic knowledge of how to control measuring instruments using a PC You can download the most recent version of these manuals from the Kikusui Electronics Corporation website http www kikusui...

Страница 3: ...tion 15 LAN connections 16 RS485 connection 17 3 Webpage 18 Home Page 18 Opening the Home page 18 Description of the Home page 19 Logging In and Logging Out 20 Logging in 20 Logging Out 21 DC Power Pa...

Страница 4: ...tions using the Internet Webpage Network communication configuration Controlling and reading the PAV output and status from a GUI screen Multi user control and security settings to eliminate dangerous...

Страница 5: ...Ethernet RJ 45 port standard 8 pin jack port for LAN The RJ 45 port is equipped with a link LED and activity LED The LAN status indicator LED on the rear panel shows the LAN communication status The...

Страница 6: ...he PAV can connect to LAN PAV identification Green blinking1 2 1 You can use the identification function by accessing the webpage from a PC or the like or by sending an SCPI command When used this LED...

Страница 7: ...utomatically assigns its own IP address and other LAN parameters To return to the default settings perform a LAN reset p 14 In addition if the PC is configured for automatic IP address assignment the...

Страница 8: ...Address MAC Address LAN Reset RS232 RS485 USB LAN 1 31 1200 SCPI IP 1 IP 4 MAC 1 MAC 6 Reset nnnn nnnn REM Parameter 1200 2400 4800 9600 19200 38400 57600 PAG Adr INtF 232 485 USb LAn SCPI PAG bAUd LA...

Страница 9: ...ace LAn 5 Press the CURRENT knob When the interface is set the display blinks and returns to the original screen Subsystem Function item Panel display item Panel display Interface INtF RS2321 1 Factor...

Страница 10: ...AV is in LAN remote mode the front panel REM LED is lit Press REM to turn off the REM LED and switch to local mode If the PAV does not switch to local mode even when you press REM perform one of the f...

Страница 11: ...nel LAN status indicator LED lights green about 10 seconds later the connection is complete In a P2P peer to peer network When the PAV rear panel LAN status indicator LED lights green about 30 to 40 s...

Страница 12: ...ss assignment DHCP auto IP and static fixed IP If you change the IP address from the front panel or by performing the procedure on the Modify page on the Configure tab of the LAN page on the webpage t...

Страница 13: ...n assign unique host names to each PAV Displaying the MAC address To display the MAC address on the front panel follow the procedure below 1 Press REM 2 Turn the VOLTAGE knob until the voltmeter shows...

Страница 14: ...about 1 second 6 Turn the POWER switch off and then back on Default LAN settings Item Default value See TCP IP mode DHCP enabled If DHCP setting fails auto IP assignment is used p 29 IP address 169 25...

Страница 15: ...RS485 link cables In a multi drop connection only a single controller e g PC can connect to the PAV with LAN at any given time Make sure it is a one to one connection IN OUT IN OUT RS485 link cable RS...

Страница 16: ...addresses of each PAV other than the master unit Setting the RS485 addresses On PAVs without LAN set RS485 addresses If PAVs without LAN are in analog remote mode switch to local mode p 10 1 Press RE...

Страница 17: ...RS485 link cable connect the J4 OUT connector of the master unit to the J4 IN connector of a slave unit 2 Turn on the POWER switch of each PAV 3 Set each PAV to RS485 remote mode 4 Set each PAV s add...

Страница 18: ...D is lit green the PAV webpages are avail able 1 Check the IP address from the front panel p 11 2 On the PC open Internet Explorer or other web browser 3 Enter the PAV s IP address or host name as sho...

Страница 19: ...85 Address This address is necessary for multi drop connection The PAV with LAN is the master unit of the multi drop connection 4 Hostname A unique name of a device on the network For details on defau...

Страница 20: ...ears 3 In the Username box enter admin 4 In the Password box enter the password The default password is blank We recommend that you set a password after logging in 5 Click Login in the Login window Th...

Страница 21: ...nge the PAV settings The webpages can be viewed by several users simultaneously but the data rate will be lower If an auto control program is in progress through a VISA or socket connection webpages c...

Страница 22: ...played in the Multi Drop RS 485 Address list box This button is available for PAV with LAN p 15 2 Multi Drop RS 485 Address This address is valid when a RS485 multi drop connection is config ured If a...

Страница 23: ...oltage and Current Limit is for setting the CC current Output is for turning the output on and off This area shows the present PAV settings To change the settings select the Check To Modify check box...

Страница 24: ...t to the maximum value that can be set Set to value The value is set to the value that you enter 2 Under Voltage Protection Enter a new value in New Setting and click Apply to change the under voltage...

Страница 25: ...2 Reset All Instruments button reset entire group Resets all the PAVs connected in the multi drop configuration 3 Save button Saves the settings of the PAV single unit selected in the Multi Drop RS48...

Страница 26: ...input text box Some commands do not have responses 2 Read System Errors Click Read Errors to display system error messages in this text box To view system error message send a SYSTem ERRor ENABle in a...

Страница 27: ...using DHCP auto IP or static fixed IP 3 Subnet Mask Displays the subnet mask of DHCP auto IP or static fixed IP 4 Default Gateway Displays the address of the network router that the PAV uses to com m...

Страница 28: ...ew address If changing the parameters results in a duplicate IP address the LAN status indicator LED and the front panel LED blink and the parameters return to their original values To stop the LED bl...

Страница 29: ...Subnet Mask Default Gate way and DNS Server parameters they are unavailable If these parameters are not assigned by the server the PAV recov ers in Auto IP mode which is described in IP address explai...

Страница 30: ...Ping Server Ping is used to check the connection to the server from a PAV con nected to the LAN through the network utility This function is disabled while the LAN Modify window on the Advanced tab of...

Страница 31: ...end a SYST COMM LAN RES command p 42 1 2 3 Users tab 6 No Item Description 1 Enter old password The default value is blank because there is no set password 2 Enter new password Enter a password using...

Страница 32: ...32 PAV LAN Help Page Click Help to open the Help page There is a link to the Kikusui website...

Страница 33: ...s with the NetBIOS protocol Can be set from the webpage Duplicate IP detection Rejects duplicate IP addresses or disconnects from the network Subnet mask Assigned by DHCP or can be set from the webpag...

Страница 34: ...ed on the front panel Multi drop address The RS485 address can be viewed on the front panel Link LED Lights when an Ethernet cable is connected to the PAV Activity LED Lights when LAN packets are dete...

Страница 35: ...ices on the network may be set to the same IP address If the PAV is in this condition an IP address cannot be assigned This may occur when the PAV is in static fixed IP address mode p 11 Perform eithe...

Страница 36: ...uc cessful If an error occurs there may be a problem with the LAN set tings of the PAV or PC or both or the PAV s ping function may be disabled p 30 If an error occurs perform a LAN reset on the PAV p...

Страница 37: ...I 11 Device_write Writes text in a measuring instrument VXI 11 Device_read Reads text from a measuring instrument VXI 11 Destroy_link Disconnects the link to a measuring instrument Starting a VISA con...

Страница 38: ...t can be configured with a management utility such as National Instruments Measurement and Automation Explorer MAX program or the Keysight Technologies I O Libraries It is also possible to configure I...

Страница 39: ...socket Controller access single or multiple clients Security settings can be applied to webpages to limit the number of PCs clients that connect to the PAV at any given time Limitation to single and...

Страница 40: ...age return is appended automatically to responses to queries If the LAN controller is set to Multiple clients from the webpage set Controller Access on the Modify page on the Configure tab of the LAN...

Страница 41: ...k your network administrator to assign a global IP address to the PAV with LAN In addi tion configure the port transfer setting of the network server s port number 80 to allow a WAN connection Socket...

Страница 42: ...15 characters in length SYST COMM LAN MAC Queries the MAC address Command SYSTem COMMunicate LAN MAC Ex SYST COMM LAN MAC Response Returns the MAC address in CRD format in response to SYST COMM LAN MA...

Страница 43: ...of the PAV with address 4 is 90 V There is no need to send the INSTrument NSELect address command again Responses to global commands When you send a global command the command is applied to all PAVs i...

Страница 44: ...rt 6 LAN reset 14 LAN status indicator LED 6 LAN Timeout 30 link LED 6 login 20 logout 21 M MAC address 13 Multi drop connection 15 Multiple clients 29 O One client only 29 P P2P peer to peer 7 11 pas...

Страница 45: ......

Страница 46: ...copy can be provided for a fee In either case please contact your Kikusui agent or distributor At that time inform your agent or distributor of the Part No written on the front cover of this manual Ev...

Отзывы: