KPM1000
Specifications
35
4
Specifications
This chapter contains the specifications and gives the dimensions of the KPM1000.
Unless specified otherwise, the specifications are for the following settings and conditions.
• The warm-up time is 30 minutes.
• After the KPM1000 has been warmed up, it must be calibrated correctly in a 23 °C ± 5 °C
environment.
Input
Display items
Item
Specification
Measurement line
Single-phase, two-wire system (measurement category: CAT II)
1
1
Applies to measurements on circuits directly connected to a low-voltage installation. This category applies
to measurements on circuits of equipment on the primary side of a transformer. Such pieces of equipment
have a power cord connected to a power outlet. Examples are household appliances and portable tools.
Voltage input terminals
Safety terminals
Current input terminals
M6 terminal block
Rated measured voltage
300 Vrms
Rated measured current
20 Arms
Maximum allowable input voltage
900 Vpk or 360 Vrms
Maximum allowable input current
120 Apk or 24 Arms
Maximum isolation voltage
300 V
Input impedance
(50/60 Hz)
Voltage input
6 M
Ω
± 10 %
Current input
2 m
Ω
or less
Line filter (LPF)
Cutoff frequency
500 Hz (can be turned on or off)
Item
Specification
Measurement items
Voltage, current, active power, apparent power, reactive power,
power factor, phase angle, frequency,
integrated current, integrated power,
positive integrated power, negative integrated power,
integration time
voltage crest factor, current crest factor,
voltage peak, current peak.
Display update interval
100 ms, 200 ms , 500 ms , 1 s , 2 s, 5 s , 10 s.
Number of displayed items
4 items simultaneously.
Display digits
Displays A and C
5.
Displays B and D
7.
Display items
Displays A and C
V, A, W, Vpk, Apk, PF, Hz.
Displays B and D
V, A, W, VA, var, Ah, Wh, [P]Wh, [n]Wh, h, [C]V, [C]A, [d].
([P] means positive, [n] means negative, [C] means crest
factor, and [d] means phase angle.)