20
Basic Operations
KPM1000
Displayed contents
Display configuration
• Sign display
If there is no sign, the value is positive. If “-” is displayed, the value is negative.
• Symbol display
If “P” is displayed, the value is the positive integrated power. If “n” is displayed, the value
is the negative integrated power.
• Numeric display
A 5- or 6-digit number and a decimal point are displayed. The number of displayed digits
is determined by the display item.
• Unit display
The unit—m (milli), k (kilo), or M (mega)—is selected automatically on the basis of the
selected range. For Vpk, the V and pk indicators light independently. In this manner, the
KPM1000 combines unit characters to display units according to the display item.
Display examples
Display item
The items that you can show on the displays are listed below.
Displays A and C
8.00000 Wh integrated power
(when the sum is negative)
99 hour, 59 minute, 59 second integration time
1.0000 voltage crest factor
-180.00 ° phase angle
8.00000 Wh integrated power (negative)
8.00000 Wh integrated power (positive)
Display item
Unit display
(center column)
Unit display
(right column)
Description
Voltage
V
-
Values that are 0.5 % of the voltage range or less
are displayed as 0.0000 (they are muted).
This is the factory default setting for display A.
Current
A
-
Values that are 0.5 % of the current range or less
are displayed as 0.0000 (they are muted).
This is the factory default setting for display C.
Active power
W
-
-
Voltage peak
V
pk
The larger of the positive and negative peak
values is displayed.
Current peak
A
pk
The larger of the positive and negative peak
values is displayed.
Power factor
-
PF
If the voltage or current is muted, “-----” is
displayed.
Frequency
-
Hz
Use the synchronization source setting (ConF.7
SynC) to select whether to display the voltage
frequency or current frequency.
If you set SynC to “oFF,” “-----” is displayed.