![Keysight U4421A Скачать руководство пользователя страница 12](http://html1.mh-extra.com/html/keysight/u4421a/u4421a_hardware-and-probing-manual_1961143012.webp)
1
U4421A Module
12
Keysight U4421A MIPI D-PHY Analyzer and Exerciser Hardware and Probing Guide
Trigger In/Out
These connectors are used to listen to an external trigger in from a device or to send an external trigger
out to another device. However, these connectors on the U4421A module are not currently functional.
Instead of using these connectors, you can use the Trigger In/Out connectors located on the Embedded
System Module (ESM) of the Keysight AXIe chassis to send or receive external trigger in and out events.
The following are some important points about the Trigger In/Out connectors:
•
The Trigger In connector of the AXIe chassis has an adjustable threshold input of +/-5V range and 250
mV minimum swing.
•
The Trigger Out connector of the AXIe chassis extends the parallel trigger bus to external instruments.
The trigger out characteristics are 3.3V CMOS, 50W line drive, and 3-state.
•
Maximum trigger input voltage should not exceed 3.3 V.
•
Trigger Out and 10 MHz Out have nominal output level of 2.0 V with 20 ns minimum pulse width.
•
Minimum Trigger In duration is 20 ns.
OOS (Out of Service)
LED
Indicates the power-ready status of the U4421A module. This LED may turn red during the
power-on-self-test phase of the chassis boot cycle. The LED then turns off when the U4421A module is in
a power-ready state. The LED turns red again only when a power fault condition occurs for the module.
• Red, steady
- If the LED remains red and does not turn off, then it indicates a power fault condition. In
such a situation, the module may require repair/service. Contact your Keysight representative to
replace or service the module.
• Off
- The module has detected no power failures and is in a power-ready status.
Component
Description
WARNING
Do not directly touch any component on the U4421A module. It may be hot.
CAUTION
Components on the U4421A module are sensitive to static electricity. Therefore,
take necessary anti-static precautions, such as wear a grounded wrist strap, to
minimize the possibility of electrostatic damage.