3
CALibration Subsystem
Remote Calibration Procedures
56
U3606A Programmer’s Reference
Remote Calibration Procedures
The CALibration commands are used to calibrate the U3606A. Please note
that the use of these commands requires a detailed knowledge of the
appropriate calibration procedures, which are described in the
U3606A
User’s and Service Guide
. Please refer to that guide before attempting to
calibrate the instrument. Improper use of the CALibration commands can
adversely affect the accuracy and reliability of the instrument.
During calibration the following instrument behavior is expected:
•
The display “CALib” in the lower secondary display starts flashing to
indicate that the calibration is in progress.
•
Successful completion of the adjustment is indicated by a short beep
and the primary display briefly showing “PASS”.
•
An adjustment failure is indicated by a long beep, the primary display
showing “FAiL” and a calibration error number appearing in the upper
secondary display. Correct the problem and repeat this procedure.
The adjustment data is stored only when all the calibration items for the
measurement selected is completed. For example, to store the adjustment
data for DC voltage measurements, you will need to complete the following
calibration items: Short, 100 mV, 1 V, –1 V, 10 V, 100 V, and 1000 V.
Before performing and adjustments, first you will need to unsecure the
instrument for calibration.
Calibration steps
Remote commands
1
Enter the calibration mode.
&
CAL:SEC:STAT OFF,
ATU3606A
2
Optional step: Change the default security
code after unsecuring the instrument for
calibration (be sure to write down the new
code).
&
CAL:SEC:CODE <new code>
3
Perform the zero and gain adjustments for
the front input terminals. See
4
Perform the output adjustments for the front
and rear output terminals. See
.
Содержание U3606A
Страница 1: ...Keysight U3606A Multimeter DC Power Supply Programmer s Reference ...
Страница 2: ......
Страница 3: ......
Страница 13: ...U3606A Programmer s Reference XI Device specific errors 341 Self test errors 342 Calibration errors 343 ...
Страница 14: ...XII U3606A Programmer s Reference ...
Страница 27: ...Introduction to SCPI 1 SCPI Status System U3606A Programmer s Reference 13 Figure 1 1 Status system diagram ...
Страница 32: ...1 Introduction to SCPI SCPI Status System 18 U3606A Programmer s Reference ...
Страница 58: ...2 CALCulate Subsystem CALCulate NULL OFFSet 44 U3606A Programmer s Reference ...
Страница 92: ...3 CALibration Subsystem Remote Calibration Procedures 78 U3606A Programmer s Reference ...
Страница 132: ...4 CONFigure Subsystem CONFigure DCYCle 118 U3606A Programmer s Reference ...
Страница 138: ...5 INITiate Subsystem INITiate CONTinuous 124 U3606A Programmer s Reference ...
Страница 174: ...7 MEMory Subsystem MEMory STATe RECall AUTO 160 U3606A Programmer s Reference ...
Страница 295: ...SOURce Subsystem 10 SOURce PROTection STATe U3606A Programmer s Reference 281 RST on page 328 ...
Страница 296: ...10 SOURce Subsystem SOURce PROTection STATe 282 U3606A Programmer s Reference ...
Страница 332: ...14 Root Commands READ 318 U3606A Programmer s Reference ...
Страница 350: ...15 IEEE 488 2 Common Commands WAI 336 U3606A Programmer s Reference ...
Страница 370: ...Command Quick Reference 356 U3606A Programmer s Reference ...